Quadrox-I Small Adult Microporous Membrane Oxygenator and 1 more

Oxygenator, Cardiopulmonary Bypass. 510(k) K093522, from Maquet Cardiopulmonary, AG (Hirrlingen, DE). Product code DTZ, class 2.

As of Nov 21, 2026, K093522 is an FDA 510(k) held by Maquet Cardiopulmonary, AG, substantially equivalent on Dec 7, 2009 after 24 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 24 days
  • Recalls: 3
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The membrane oxygenator QUADROX-i Small Adult is intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. The blood flow rate is defined from 0.5 – 5 l/min. Within the specified flow rate range, the device oxygenates the blood, eliminates carbon dioxide and regulates blood temperature.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K093522, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Feb 23, 2016Z-1371-2016Maquet Cardiopulmonary (MCP) is conducting a voluntary recall based on the potential risk of exposure to endotoxin for various products.Terminated
Aug 15, 2013Z-2262-2013The manufacturer has conducted a review of the results of product testing performed during the last twelve months and has identified specific lots of product which may not meet specifications for bond strength on the…Terminated
Jun 12, 2012Z-0160-2013It has come to firm's attention that, in some rare cases, the blood outlet connector on some models of MAQUET Oxygenators may detach from the oxygenator. This event has occurred in about 0.01% (1:10,000) of units…Terminated

Other DTZ clearances

The newest 510(k)s for the same product code; their median review was 269 days; see every DTZ device

Record

Type
Special
Received
Nov 13, 2009
Review panel
Cardiovascular
Regulation
21 CFR 870.4350

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files