Acuson S2000 Diagnostic Ultrasound System

System, Imaging, Pulsed Doppler, Ultrasonic. 510(k) K082142, from Siemens Medical Solutions USA, Inc., Ultrasound DI (Mountainview, CA). Product code IYN, class 2.

As of Nov 27, 2026, K082142 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., Ultrasound DI, substantially equivalent on Nov 13, 2008 after 106 days, with 8 recalls tied to it in FDA's recall database.

  • Review time: 106 days
  • Recalls: 8

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The S2000™ ultrasound imaging systems are intended for the following applications: Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.

Recalls

Recalls FDA ties to K082142, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 15, 2011Z-2051-2011The Firm discovered a software error on the Accuson S2000 ultrasound systems affecting the measurement results in the patient report.Terminated
Aug 17, 2010Z-1837-2011There is an error in the estimation of mean pressure gradient in the Cardiac measurement package. The currently implemented calculation will result in an underestimation of the mean pressure gradient (PGmean) when…Terminated
Oct 2, 2009Z-1006-2011, Z-1005-2011, Z-1007-2011ECG waveform does not accurately align with 2D, M-Mode or Spectral Images.Terminated
Aug 28, 2009Z-1392-2011If the transducer is not disconnected from the system after an automatic shutdown due to thermal limits being exceeded, then the next time it is used, the thermal safety guards do not take effect. Normally, the thermal…Terminated
Apr 7, 2009Z-1884-2011Device function failure--When performing a TEE exam with the transducer, CW (Continuous Wave) mode is not functional. Due to work flow, this may not be apparent until the transducer is inserted into the patient's…Terminated
Jan 14, 2009Z-1900-2011The product has a software problem in which previous patient measurement data gets associated with another patient's image.Terminated

Other IYN clearances

The newest 510(k)s for the same product code; their median review was 134 days; see every IYN device

Record

Type
Traditional
Received
Jul 30, 2008
Review panel
Radiology
Regulation
21 CFR 892.1550

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files