Aw Server
System, Image Processing, Radiological. 510(k) K081985, from Ge Medical Systems, LLC (Waukesha, WI). Product code LLZ, class 2.
As of Nov 18, 2026, K081985 is an FDA 510(k) held by Ge Medical Systems, LLC, substantially equivalent on Sep 25, 2008 after 73 days, with 5 recalls tied to it in FDA's recall database.
- Review time: 73 days
- Recalls: 5
- Cited as a predicate by: 5 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDF2/3 GE Medical Systems 6. Comparison with Predicate Device The functional features of AW Server software package are substantially equivalent to that of the following devices: Device Name FDA Clearance Number Advantage Workstation 4.3 K052995 AquariusNET Server K012086 7.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Advantage Workstation 4.3.K052995 · GE Healthcare · Nov 8, 2005
- Aquariusnet ServerK012086 · Terarecon,Inc. · Sep 13, 2001
Cited as a predicate
5 later 510(k) summaries name this device as a predicate- FlightPlan for EmbolizationK193261 · GE Medical Systems SCS · Jan 24, 2020
- IntelliSpace Portal PlatformK162025 · Philips Medical Systems Nederlands B.V. · Oct 18, 2016
- Customer Remote Console (CRC)K150193 · Ge Medical Systems, LLC · Mar 24, 2015
- Myrian 1.11K113620 · Intrasense · Jul 12, 2012
- Onco Trac Model 1, 2, 4K111642 · Translational Sciences Corporation · Nov 18, 2011
Recalls
Recalls FDA ties to K081985, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| May 8, 2026 | Z-2553-2026GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches an interactive application (e.g.… | Open |
| Sep 5, 2025 | Z-0068-2026Firm has identified a security vulnerability in AW Server products. If exploited, a malicious actor could compromise the confidentiality, integrity, and availability of patient data. | Open |
| Sep 19, 2013 | Z-0270-2014GE Healthcare has recently become aware of a potential safety issue due to the image orientation within GSI viewer on AW products. The GSI Viewer application running on AW/Volume Share or on the AW Server when making a… | Terminated |
| Jun 28, 2012 | Z-1137-2013GE Healthcare has recently become aware of a potential safety issue with respect to the Results Viewer in the AW Server. After starting the Results Viewer with the selected series of a patient, the pixel and certain… | Terminated |
| Nov 17, 2011 | Z-0486-20122 Unrelated Software Issues: An incorrect patient name maybe displayed when switching between two Volume Viewer tabs with different patients. Also, if a second patient exam is selected while the application is starting,… | Terminated |
Other LLZ clearances
The newest 510(k)s for the same product code; their median review was 145 days; see every LLZ device- 3D Anatomical ModelK262132 · The Jacobs Institute · Nov 5, 2026
- LiLi DICOM Viewer (LiLi )K260454 · Elevate Imaging, Inc. · Nov 4, 2026
- Vitrea Advanced Visualization, Version 7.16K263648 · Canon Medical Informatics, Inc. · Oct 29, 2026
- MIM – KineticIDK261763 · Mim Software, Inc. · Oct 26, 2026
- PV-iCurve ProK261531 · Perception Vision Medical Technology Co., Ltd. · Oct 21, 2026
- Eden PACSK261788 · Higia., Inc. d/b/a Eden · Oct 21, 2026
- Syngo Carbon Space (VA4)K263048 · Siemens Healthineers AG · Sep 29, 2026
- Syneris PACS v.1.0.1.0K261874 · Syneris · Sep 23, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Jul 14, 2008
- Review panel
- Radiology
- Regulation
- 21 CFR 892.2050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files