Advantage Workstation 4.3.
System, Image Processing, Radiological. 510(k) K052995, from GE Healthcare (Waukesha, WI). Product code LLZ, class 2.
As of Nov 18, 2026, K052995 is an FDA 510(k) held by GE Healthcare, substantially equivalent on Nov 8, 2005 after 14 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 14 days
- Recalls: 4
- Cited as a predicate by: 9 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFAdvantage Workstation 4.3 is a review station, which allows easy selection, review, processing and filming of multi-modality DICOM images from a variety of diagnostic imaging systems. When interpreted by a trained physician, filmed or displayed images on the AW monitor may be used as a basis for diagnosis, except in the case of mammography images.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Advantage Workstation 4.1K020483 · GE Medical Systems · Feb 28, 2002
- Volume Viewer PlusK041521 · General Electric Medical Systems · Jun 22, 2004
Cited as a predicate
9 later 510(k) summaries name this device as a predicate- Flightplan for LiverK121200 · GE Healthcare · Nov 2, 2012
- Aw Volumeshare 5 With Angioviz OptionK110834 · GE Healthcare · Apr 26, 2011
- Innova Vision, Model S18751VN, Innova Trackvision, Model S18751VT, Innova Epvision, Model S18751EPK092639 · GE Medical Systems SCS · Dec 2, 2009
- Myrian 1.4K091001 · Intrasense · Jun 29, 2009
- UrocareK090552 · Neusoft Medical Systems Co., Ltd. · Mar 16, 2009
- Aw ServerK081985 · Ge Medical Systems, LLC · Sep 25, 2008
- MyrianK071000 · Intrasense · May 14, 2007
- Fullcard AnalysisK061587 · GE Healthcare · Jun 23, 2006
- Aze VirtualplaceK060453 · Aze, Ltd. · Mar 9, 2006
Recalls
Recalls FDA ties to K052995, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Sep 19, 2013 | Z-0270-2014GE Healthcare has recently become aware of a potential safety issue due to the image orientation within GSI viewer on AW products. The GSI Viewer application running on AW/Volume Share or on the AW Server when making a… | Terminated |
| May 24, 2012 | Z-2203-2012GE Healthcare is informing you of incorrect measurements when using the Aorta analysis protocol or customized protocols. Until a new software version is installed they are asking you not to use the aorta analysis… | Terminated |
| May 31, 2011 | Z-3012-2011GE Healthcare has become aware that patient mismatch could occur on some Seno Advantage 2.1 and 2.2 Workstations with RIS CCOW integration. This issue may impact patient safety. When Seno Advantage workstation is… | Terminated |
| Sep 10, 2007 | Z-0265-2008Aortic Length over-estimated: incorrect measurements when using the Aorta analysis protocol or customized protocols. Until a new software version is installed the aorta analysis protocol or the customized protocols for… | Terminated |
Other LLZ clearances
The newest 510(k)s for the same product code; their median review was 145 days; see every LLZ device- 3D Anatomical ModelK262132 · The Jacobs Institute · Nov 5, 2026
- LiLi DICOM Viewer (LiLi )K260454 · Elevate Imaging, Inc. · Nov 4, 2026
- Vitrea Advanced Visualization, Version 7.16K263648 · Canon Medical Informatics, Inc. · Oct 29, 2026
- MIM – KineticIDK261763 · Mim Software, Inc. · Oct 26, 2026
- PV-iCurve ProK261531 · Perception Vision Medical Technology Co., Ltd. · Oct 21, 2026
- Eden PACSK261788 · Higia., Inc. d/b/a Eden · Oct 21, 2026
- Syngo Carbon Space (VA4)K263048 · Siemens Healthineers AG · Sep 29, 2026
- Syneris PACS v.1.0.1.0K261874 · Syneris · Sep 23, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Oct 25, 2005
- Review panel
- Radiology
- Regulation
- 21 CFR 892.2050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files