Ventrio Hernia Patch
Mesh, Surgical, Polymeric. 510(k) K081777, from C.R. Bard, Inc. (Warwick, RI). Product code FTL, class 2.
As of Nov 17, 2026, K081777 is an FDA 510(k) held by C.R. Bard, Inc., substantially equivalent on Sep 29, 2008 after 98 days, with 4 recalls tied to it in FDA's recall database.
- Substantially equivalent: Sep 29, 2008
- Review time: 98 days
- Recalls: 4
- Cited as a predicate by: 7 devices
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Bard Extra Large Composix Kugel Hernia PatchK061314 · Davol Inc., Sub. C. R. Bard, Inc. · Jun 2, 2006
- Proceed Ventral PatchK061533 · ETHICON, Inc. · Dec 11, 2006
- Gore Infinit MeshK081069 · W. L. Gore & Associates, Inc. · May 30, 2008
- Dualmesh Emerge & Emerge Plus BiomaterialK022782 · W. L. Gore & Associates, Inc. · Sep 10, 2002
Cited as a Predicate
7 later 510(k) summaries name this device as a predicate- ONFLEX MeshK142711 · C.R. Bard, Inc. · Mar 20, 2015
- Modified ONFLEX MeshK142706 · C.R. Bard, Inc. · Mar 25, 2015
- Bard Ventralex Hernia PatchK132441 · C.R. Bard, Inc. · Dec 13, 2013
- Ventrio Light Hernia Patch With Trm Antimicrobial CoatingK113229 · C.R. Bard, Inc. · Jul 20, 2012
- Ventralex St Hernia PatchK101928 · C.R. Bard, Inc. · Mar 23, 2011
- Ventrio St Hernia PatchK101920 · C.R. Bard, Inc. · Mar 23, 2011
- Ventrio Hernia PatchK100229 · C.R. Bard, Inc. · Apr 21, 2010
Recalls
Recalls FDA ties to K081777, newest first; each opens FDA's recall record| Oct 15, 2010 | Z-0423-2011Mislabeled: Hernia Patch may be a different size than on the product label | Terminated |
| Oct 15, 2010 | Z-0424-2011Mislabeled: Hernia Patch may be a different size than on the product label | Terminated |
| Oct 15, 2010 | Z-0422-2011Mislabeled: Hernia Patch may be a different size than on the product label | Terminated |
| Apr 8, 2010 | Z-1536-2010Mislabeled: Product labeled as Small Oval (3.1" x 4.7") may contain Small Circle (3.0") | Terminated |
Other FTL Clearances
The newest 510(k)s for the same product code; their median review was 85 days; see every FTL device- Liora Monofilament ScaffoldK260474 · TELA Bio, Inc. · Sep 28, 2026
- ProGrip advanced self-gripping polypropylene mesh (ADG1510, ADG2015, ADG3020, ADG4030)K253956 · Sofradim Production · May 21, 2026
- Parietene Flat Sheet MeshK253125 · Medtronic - Sofradim Production · Oct 24, 2025
- Onflex MeshK251955 · Davol Inc., Subsidiary of C. R. Bard, Inc. · Jul 24, 2025
- Bard® Mesh; Bard® Mesh Pre-ShapedK251557 · Davol, Inc. · Jun 12, 2025
- Parietene Macroporous Mesh (PPM5050 )K250869 · Sofradim Production · Apr 23, 2025
- Bard Soft Mesh; Bard Soft Mesh Pre-ShapedK250098 · Davol, Inc. · Apr 9, 2025
- ProGrip Self-Gripping Polypropylene MeshK243315 · Sofradim Production · Jan 17, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jun 23, 2008
- Review panel
- General, Plastic Surgery
- Regulation
- 21 CFR 878.3300
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files