Plum A+ Hyperbaric Infusion System With Hospira Mednet Software

Pump, Infusion. 510(k) K081412, from Hospira, Inc. (Lake Forest, IL). Product code FRN, class 2.

As of Nov 20, 2026, K081412 is an FDA 510(k) held by Hospira, Inc., substantially equivalent on Dec 17, 2008 after 211 days, with 17 recalls tied to it in FDA's recall database.

  • Review time: 211 days
  • Recalls: 17

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Plum A+ Hyperbaric Infusion System with Hospira MedNet Software is indicated for use in parenteral, enteral and epidural therapies and the administration of whole blood and blood products. Additionally, the Plum A+ Hyperbaric Infusion System with Hospira MedNet Software can be used under hyperbaric conditions in multiplace and monoplace (Class A and B) chambers.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K081412, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 2, 2014Z-1074-2015One lot of alarm assemblies used in Plum A+ and Plum A+3 infusion pumps may fail to sound at all volume levels.Terminated
Feb 6, 2013Z-2729-2016, Z-2725-2016, Z-2726-2016, Z-2724-2016, Z-2727-2016, Z-2730-2016, Z-2722-2016, Z-2728-2016, Z-2723-2016The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and delay or interruption of therapy.Terminated
Feb 1, 2013Z-0283-2014Plum A+ infusers have the potential for the distal (occlusion) press sensor pin to break. The distal pressure sensor pin is a part of the overall subsystem that measures the pressure within the distal line of the…Terminated
Dec 7, 2012Z-0626-2013The door roller assembly on the Plum A+/A+3 infuser family of infusers has the potential to break which can lead to possible unrestricted flow and/or over-delivery during the removal of the IV administration set's…Terminated
Oct 26, 2012Z-0482-2013The volume control knob on some Plum A+ single channel infusers (located on the back of the infuser) may not function as described in the System Operating Manual - the direction for Loud and Quiet may be reversed. The…Terminated
Feb 5, 2012Z-0203-2014Hospira has received reports of distal pressure sensor calibration drift on Plum A+ Infusers. The following error codes may be displayed during setup or infusion signifying possible calibration drift: E180/N180,…Terminated
Sep 16, 2011Z-0011-2012Hospira has received reports of continuous recycling and/or rebooting of Plum A+ infusion pumps when the backlight intensity and/or display contrast settings for the LED display has been adjusted from the original…Terminated
Sep 16, 2011Z-3286-2011Hospira has received reports of incorrect seating of the regulator closer on the Plum A+ Infusion Pumps. If the regulator closer does not close the flow regulator actuator when the pump door is opened, and if the…Terminated
Feb 14, 2011Z-1646-2011Hospira has received customer reports of the Plum A+ infusion pumps with no audible alarm conditions, which have been associated with failure of the piezoelectric assembly due to component quality issues.Terminated

Other FRN clearances

The newest 510(k)s for the same product code; their median review was 242 days; see every FRN device

Record

Type
Traditional, third-party review
Received
May 20, 2008
Review panel
General Hospital
Regulation
21 CFR 880.5725

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files