Arca
Skip to content
Tools
Tools

4D Integrated Treatment Console With Patient Accessory Verification System (Pavs)

Accelerator, Linear, Medical. 510(k) K081036, from Varian Medical Systems, Inc. (Palo Alto, CA). Product code IYE, class 2.

As of Nov 20, 2026, K081036 is an FDA 510(k) held by Varian Medical Systems, Inc., substantially equivalent on May 9, 2008 after 28 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 28 days
  • Recalls: 3
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The 4DITC function is designed to assist the operator of a radiation therapy device in providing accurate treatment setups for each patient by monitoring setup parameters and preventing the radiation therapy device from commencing irradiation when any parameter is out of conformance with the treatment plan.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K081036, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jun 27, 2011Z-3100-2011In certain configurations, the 4DITC can interpret a RapidArc or VMAT treatment field as 3D Conformal Arc Treatment Plan. If a RapidArc or VMAT treatment field is incorrectly interpreted as Conformal Arc, the field will…Terminated
Jun 23, 2011Z-3098-2011Varian has scheduled for an upgrade of the device from current version to 10.2. There are circumstances in which the 4DITC may be involved in a failure to correctly record treatment history information.Terminated
Dec 14, 2010Z-0932-2011Multiple isocenter plans will load on the 4DITC and are available for imaging and treatment, but the 4DITC software does not recognize that mulitplex isocenter is present.Terminated

Other IYE clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every IYE device

Record

Type
Special
Received
Apr 11, 2008
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files