Arca
Skip to content
Tools
Tools

4D Integrated Treatment Console

Accelerator, Linear, Medical. 510(k) K050479, from Varian Medical Systems (Palo Alto, CA). Product code IYE, class 2.

As of Nov 20, 2026, K050479 is an FDA 510(k) held by Varian Medical Systems, substantially equivalent on Mar 24, 2005 after 28 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 28 days
  • Recalls: 2
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The 4D Integrated Treatment Console is designed to assist the operator of a radiation therapy device in providing accurate treatment set-ups for each patient by monitoring set up parameters and preventing the radiation therapy device from commencing irradiation while any parameter is out of conformance with the treatment plan.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • Varis 1.4GK001643 · Varian Medical Systems, Inc. · Jun 23, 2000

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K050479, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Sep 21, 2006Z-0304-2007Selecting an empty space next to the ''Override'' or ''Acquire Actual'' buttons may result in mistreatment to the patient.Terminated
Sep 20, 2005Z-1375-06A software anomaly may occur which can lead to patient treatment with the wrong field. The anomaly is reported only to occur when this software version (6.6.5022) is used with Elekta or Siemens linear accelerators .Terminated

Other IYE clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every IYE device

Record

Type
Special
Received
Feb 24, 2005
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files