Dimension Exl With Lm Clinical Chemistry System, Dimension FT4L Flex Reagent Cartridge, And Loci Thyroid Calibrator, Mod

Radioimmunoassay, Free Thyroxine. 510(k) K073604, from Dade Behring, Inc. (Newark, DE). Product code CEC, class 2.

As of Nov 18, 2026, K073604 is an FDA 510(k) held by Dade Behring, Inc., substantially equivalent on Mar 13, 2008 after 83 days, with 7 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Mar 13, 2008
  • Review time: 83 days
  • Recalls: 7
  • Cited as a predicate by: 4 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The FT4L method is an in vitro diagnostic test for the quantitative measurement of Free Thyroxine in human serum and plasma on the Dimension® EXL™ with LM system. Measurements of free thyroxine are used in the diagnosis and monitoring of thyroid disease. The LOCI® Thyroid Calibrator is an in vitro diagnostic product for the calibration of the FT4L method on the Dimension® EXL™ with LM system.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K073604, newest first; each opens FDA's recall record
Apr 2, 2015Z-1615-2015Certain Dimension Integrated Chemistry Systems Instrument Power Configurations may have been installed with a non-standard instrument power plug.The use of a non-standard plug configuration as described may lead to a…Terminated
May 27, 2014Z-2133-2014There is a misassembly issue with the photometer installed on Dimension Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that a limited number of…Terminated
Feb 27, 2013Z-1047-2013Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined…Terminated
Feb 27, 2013Z-1048-2013Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined…Terminated
Feb 27, 2013Z-1046-2013Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined…Terminated
Apr 12, 2010Z-0458-2011Some LOCI cartridges have sub-optimal seal of the lidstock that can cause rupture of the package at high elevations, causing the cartridge to leak. Use of leaking cartridges may result in inaccurate test results.Terminated
Sep 30, 2008Z-0581-2009Error messages, "Wash Aspirator Failure" and "Wash station unable to recover" are not generated when one or both HM wash probes are totally occluded.Terminated

Other CEC Clearances

The newest 510(k)s for the same product code; see every CEC device

Record

Type
Traditional
Received
Dec 21, 2007
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1695

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files