Pressure Injectable Picc
Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days. 510(k) K073451, from Arrow International, Inc. (Reading, PA). Product code LJS, class 2.
As of Nov 19, 2026, K073451 is an FDA 510(k) held by Arrow International, Inc., substantially equivalent on Jan 15, 2008 after 36 days, with 6 recalls tied to it in FDA's recall database.
- Review time: 36 days
- Recalls: 6
- Cited as a predicate by: 2 devices
The PressureInjectable PICC is indicated for short or long term peripheral access to the central venous system for intravenous therapy, blood sampling, infusion and power injection of contrast media. The maximum pressure of power injector equipment used with the pressure injectable PICC may not exceed 300 psi.
- Arrow Pressure Injectable PiccK061289 · Arrow International, Inc. · Aug 24, 2006
- Arrowadvantages Pressure Injectable Peripherally Inscrted Central Catheter KitsK113277 · Arrow International, Inc. · Jun 21, 2012
- Pressure Injfectable PiccK080604 · Arrow International, Inc. · Jun 27, 2008
| Aug 10, 2023 | Z-2590-2023This recall has been initiated due to reports that the product code and product name were incorrectly listed on the lidstock of affected kits. | Open |
| Sep 18, 2015 | Z-0127-2016Labeling inaccuracy in the Instructions for Use (IFU) and Patient Card. The IFU and Patient Card with these products state the PICC can withstand up to ten (10) pressure injections, however, the PICC has only been… | Terminated |
| Jul 30, 2015 | Z-2462-2015the catheter peel-away component hub tabs may prematurely detach. One injury reported. | Terminated |
| Jul 30, 2015 | Z-2489-2015saline leakage from unsecured connections between the Touhy Borst adaptor and the hub of the catheter. No injuries reported. | Terminated |
| Aug 6, 2014 | Z-0066-2015The component supplier recalled their 0.9% Sodium Chloride Injection USP BD Posiflush SF Saline Flush Syringe 10mL because some unit packages may exhibit open seals which impact package integrity and potentially product… | Terminated |
| Mar 29, 2010 | Z-1682-2010Tight fit in sheath. Arrow has received complaints of the PICC catheter fitting too tightly in the "kit supplied" peelable sheath. In some cases, clinicians have been unable to insert the catheter through the peelable… | Terminated |
- Pro-Line XL Power Injectable Central Venous Catheter and Pro-Line XL TunnelerK260080 · Medical Components, Inc. · Oct 9, 2026
- Neonav ECG Tip Location SystemK260929 · Navi Medical Technologies · Apr 17, 2026
- BioFlo Hybrid PICC with ENDEXO Technology and PASV Valve Technology; BioFlo PICC with ENDEXO Technology and PASV Valve…K252373 · Spectrum Vascular · Dec 16, 2025
- PM2™ System with ECGuide™ ConnectorK252792 · Piccolo Medical, Inc. · Nov 17, 2025
- HydroPICC 4Fr Single Lumen Marked catheter, 130 cm guidewire - Basic Kit (70001301); HydroPICC 4Fr Single Lumen Marked…K243458 · Access Vascular, Inc. · Aug 1, 2025
- Dual Lumen 5Fr HydroPICC Catheter (PICC-251CM) (PICC-251CM); Dual Lumen 5Fr HydroPICC Catheter (PICC-252CM) (PICC-252CM)K251212 · Access Vascular, Inc. · Jul 1, 2025
- PM2+ System and SmartPICC StyletK243484 · Piccolo Medical, Inc. · May 29, 2025
- pilot TLSK241587 · Vygon Corporation · Feb 26, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Dec 10, 2007
- Review panel
- General Hospital
- Regulation
- 21 CFR 880.5970
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files