Modification To Avea Ventilator
Ventilator, Continuous, Facility Use. 510(k) K073069, from Viasys Respiratory Care, Inc. (Palm Spring, CA). Product code CBK, class 2.
As of Nov 18, 2026, K073069 is an FDA 510(k) held by Viasys Respiratory Care, Inc., substantially equivalent on Jan 25, 2008 after 87 days, with 14 recalls tied to it in FDA's recall database.
- Review time: 87 days
- Recalls: 14
- Cited as a predicate by: 2 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe AVEA is intended to provide continuous respiratory support in an institutional health care environment (e.g. hospitals). It may be used on adult, pediatric, and neonatal patients. It should only be operated by properly trained clinical personnel, under the direction of a physician.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Cosmo Plus With Nico, Model 8200K982499 · Novametrix Medical Systems, Inc. · Oct 16, 1998
- Servo Ventilator System, Model 64 87 800 E407EK022132 · Siemens Medical Solutions USA, Inc. · Sep 19, 2002
- Co 2 Smo PlusK963380 · Novametrix Medical Systems, Inc. · Jan 22, 1997
- AveaK013642 · Bird Products Corp. · May 22, 2002
- Bird Avea VentilatorK022674 · Bird Products Corp. · Nov 6, 2002
Cited as a predicate
2 later 510(k) summaries name this device as a predicate- Avea VentilatorK103211 · Care Fusion · May 12, 2011
- Vela VentilatorK093094 · Cardinal Health 207, Inc. · May 11, 2010
Recalls
Recalls FDA ties to K073069, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| May 17, 2016 | Z-1899-2016, Z-1888-2016, Z-1894-2016, Z-1892-2016, Z-1890-2016, Z-1891-2016, Z-1898-2016, Z-1897-2016, Z-1893-2016, Z-1896-2016, Z-1889-2016, Z-1895-2016CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to… | Terminated |
| Sep 5, 2013 | Z-2273-2013CareFusion has identified a potential risk associated with AVEA ventilators when used at higher altitudes. A proactive complaint review identified an error in the barometric pressure compensation. CareFusion is… | Terminated |
| Sep 8, 2011 | Z-0510-2012The recall was initiated because Carefusion has identified a potential risk associated with certain AVEA ventilators and affected replacement parts manufactured between March 1, 2009 and June 30, 2011. Carefusion is… | Terminated |
Other CBK clearances
The newest 510(k)s for the same product code; their median review was 235 days; see every CBK device- Evita (V800); Evita (V600)K253647 · Draegerwerk AG & CO Kgaa · May 7, 2026
- Inbentus VersatileK252663 · Iconic Solutions by Murcia, SL · Apr 14, 2026
- VPAP Pediatric Face MaskK251888 · Hsiner Co., Ltd. · Mar 12, 2026
- Vivo 45 LSK232365 · Oconnell Regulatory Consultants, Inc. · Dec 19, 2025
- F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size A (ONIV117A); F&P OptiNIV…K252173 · Fisher & Paykel Healthcare, Ltd. · Sep 10, 2025
- SV600, SV800 VentilatorK243767 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Aug 28, 2025
- AVEA disposable expiratory filter/water trap (11790)K251133 · Telesair, Inc. · May 8, 2025
- AF531 Oro-Nasal SE Face MaskK243394 · Respironics, Inc. · Dec 30, 2024
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Oct 30, 2007
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.5895
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files