Avea
Ventilator, Continuous, Facility Use. 510(k) K013642, from Bird Products Corp. (Palm Spring, CA). Product code CBK, class 2.
As of Nov 19, 2026, K013642 is an FDA 510(k) held by Bird Products Corp., substantially equivalent on May 22, 2002 after 198 days, with 19 recalls tied to it in FDA's recall database.
- Review time: 198 days
- Recalls: 19
- Cited as a predicate by: 8 devices
- Nellcor Puritan Bennett, 840 Ventilator System (Model 840)K970460 · Puritan Bennett Corp. · Apr 30, 1998
- Evita 4K961687 · Drager, Inc. · Dec 13, 1996
- Servo Ventilator 300K902859 · Siemens Life Support Systems · Apr 26, 1991
- Bear Cub 750 PsvK992788 · Bear Medical Systems, Inc. · Sep 16, 1999
- Bear 1000, Veolar W Leonardo Graphics, Bicore CP-2000, Bicore CP-100, Hi-Lo Jet Et TubeK983981 · Bear Medical Systems, Inc. · Nov 20, 1998
- V200 Ventilator With Aprv Mode, Esprit Ventilator With Aprv ModeK110083 · Respironics California, Inc. · Jan 13, 2012
- Avea VentilatorK103211 · Care Fusion · May 12, 2011
- Avea VentilatorK081837 · Cardinal Health 207, Inc. · Aug 13, 2008
- Modification To Avea VentilatorK073069 · Viasys Respiratory Care, Inc. · Jan 25, 2008
- Servo-I Ventilator System, Model 64 87 800 E407E; Nava OptionK062793 · Maquet Critical Care AB · Feb 7, 2007
- Avea VentilatorK062093 · Viasys Respiratory Care, Inc. · Sep 20, 2006
- Bird Tbird Vela VentilatorK032451 · Bird Products Corp. · Mar 17, 2004
- Bird Avea VentilatorK022674 · Bird Products Corp. · Nov 6, 2002
| May 17, 2016 | Z-1899-2016, Z-1888-2016, Z-1894-2016, Z-1892-2016, Z-1890-2016, Z-1891-2016, Z-1898-2016, Z-1897-2016, Z-1893-2016, Z-1896-2016, Z-1889-2016, Z-1895-2016CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to… | Terminated |
| Sep 5, 2013 | Z-2273-2013CareFusion has identified a potential risk associated with AVEA ventilators when used at higher altitudes. A proactive complaint review identified an error in the barometric pressure compensation. CareFusion is… | Terminated |
| Sep 8, 2011 | Z-0510-2012The recall was initiated because Carefusion has identified a potential risk associated with certain AVEA ventilators and affected replacement parts manufactured between March 1, 2009 and June 30, 2011. Carefusion is… | Terminated |
| Dec 17, 2007 | Z-0986-2008, Z-0982-2008, Z-0984-2008, Z-0983-2008, Z-0985-2008Alarm Failure -- There is a potential of alarms not being activated subsequent to specific electronic faults that may cause INOP condition. | Terminated |
- Evita (V800); Evita (V600)K253647 · Draegerwerk AG & CO Kgaa · May 7, 2026
- Inbentus VersatileK252663 · Iconic Solutions by Murcia, SL · Apr 14, 2026
- VPAP Pediatric Face MaskK251888 · Hsiner Co., Ltd. · Mar 12, 2026
- Vivo 45 LSK232365 · Oconnell Regulatory Consultants, Inc. · Dec 19, 2025
- F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size A (ONIV117A); F&P OptiNIV…K252173 · Fisher & Paykel Healthcare, Ltd. · Sep 10, 2025
- SV600, SV800 VentilatorK243767 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Aug 28, 2025
- AVEA disposable expiratory filter/water trap (11790)K251133 · Telesair, Inc. · May 8, 2025
- AF531 Oro-Nasal SE Face MaskK243394 · Respironics, Inc. · Dec 30, 2024
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Nov 5, 2001
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.5895
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files