Eclipse Treatment Planning System
System, Planning, Radiation Therapy Treatment. 510(k) K073020, from Varian Medical Systems, Inc. (Palo Alto, CA). Product code MUJ, class 2.
As of Nov 19, 2026, K073020 is an FDA 510(k) held by Varian Medical Systems, Inc., substantially equivalent on Dec 19, 2007 after 54 days, with 6 recalls tied to it in FDA's recall database.
- Review time: 54 days
- Recalls: 6
- Cited as a predicate by: 3 devices
The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments.
- Modification To Eclipse Treatment Planning SystemK071873 · Varian Medical Systems, Inc. · Aug 9, 2007
- Eclipse 6.5K010975 · Varian Medical Systems, Inc. · May 2, 2001
- Eclipse, Helios OptionK021268 · Varian Medical Systems, Inc. · May 22, 2002
- Prowess Panther Proarc, Model Version 5.0K101076 · Prowess, Inc. · Jul 27, 2010
- Modification To Eclipse Treatment Planning SystemK091492 · Varian Medical Systems, Inc. · Jun 18, 2009
- Monaco Rtp System, Vmat OptionK091179 · Computerized Medical Systems, Inc. · Jul 16, 2009
| Jul 7, 2011 | Z-3022-2011An anomaly has been identified in the 4D Structure Merge option of certain versions of Varian's Eclipse Treatment Planning System. The treatment field energy may change from the current selection to a previous selection… | Terminated |
| Jun 30, 2011 | Z-3002-2011In affected Eclipse versions, the treatment field energy may change from the current selection to a previous selection without use intervention and without invalidating the MU or calculated dose distribution. | Terminated |
| May 13, 2011 | Z-2780-2011In affected Eclipse versions, user editing of the MLC Leaf positions may not automatically invalidate calculated dose distributions. | Terminated |
| May 3, 2011 | Z-2702-2011In affected Analytical Anisotropy Algorithm versions when modeling the second source, the collimator jaw positions are read from the first control point of the field. This can cause the delivered dose to be higher than… | Terminated |
| Apr 4, 2011 | Z-2220-2011An anomaly in Eclipse Treatment Planning System, versions 8.0 to 8.6, whereby Actual SSD (source surface distance) value may not correspond to the Planned SSD value after Planning Approval. | Terminated |
| Dec 9, 2009 | Z-1412-2010Potential for misdelivery of treatment-- When preparing a treatment plan involving a split, the device may calculate excessive or insufficient monitor units for the treatment plan. | Terminated |
- RT Elements (4.6)K261655 · Brainlab SE · Oct 5, 2026
- Virtual Cone (1.0)K260750 · Varian Medical Systems, Inc. · Jul 8, 2026
- RadOncAIK253050 · Informai · Jun 18, 2026
- Multimodality Simulation Workspace (MM Sim) (v1.0.0)K260207 · Philips Medical Systems Nederland B.V. · May 27, 2026
- eXaSkin PlusK252784 · Anatomical Geometry S.L. · May 22, 2026
- ART.1-USK253078 · Seetreat Pty, Ltd. · May 14, 2026
- syngo.via RT Image Suite VC10K252304 · Varian Medical Systems, Inc. · Mar 18, 2026
- Monaco RTP System (6.3)K252002 · Elekta Solutions AB · Feb 19, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Oct 26, 2007
- Review panel
- Radiology
- Regulation
- 21 CFR 892.5050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files