Arca
Skip to content
Tools
Tools

Modification To Eclipse Treatment Planning System

System, Planning, Radiation Therapy Treatment. 510(k) K071873, from Varian Medical Systems, Inc. (Palo Alto, CA). Product code MUJ, class 2.

As of Nov 19, 2026, K071873 is an FDA 510(k) held by Varian Medical Systems, Inc., substantially equivalent on Aug 9, 2007 after 34 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 34 days
  • Recalls: 5
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K071873, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 13, 2010Z-0455-2011Incorrect treatment plan or record: A software design deficiency has the potential for corrupted treatment plan records or treatment records in cases where multiple users editing a record simultaneously can result in a…Terminated
Jul 17, 2009Z-2231-2009Software Issue/Mistreatment Potential-- Due to software issues, dose distribution may be incorrect and unintended radiation doses may be delivered.Terminated
Oct 2, 2008Z-0855-2009When using the multiple static segment option, the device may alter dose delivery.Terminated
Feb 13, 2008Z-1612-2008A software error causes the wedge accessory calculation to be ignored in the radiotherapy treatment plan.Terminated
Aug 17, 10Z-0911-2011Possible incorrect dose calculation: An anomaly in certain Eclipse versions where, under certain specific sequences of user actions, the dose distribution may fail to be re-calculated even though changes have been made…Terminated

Other MUJ clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every MUJ device

Record

Type
Special
Received
Jul 6, 2007
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files