Arca
Skip to content
Tools
Tools

Infinity Delta/Delta Xl/ Kappa/ Gamma X Xl/ Vista Xl/ SC7000/SC8000/ SC9000XL

Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms). 510(k) K070566, from Draeger Medical Systems, Inc. (Andover, MA). Product code MHX, class 2.

As of Nov 19, 2026, K070566 is an FDA 510(k) held by Draeger Medical Systems, Inc., substantially equivalent on Jun 4, 2007 after 96 days, with 6 recalls tied to it in FDA's recall database.

  • Review time: 96 days
  • Recalls: 6
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Scio gas module samples breathing gases from adults and pediatrics. The gas module continuously measure the content of CO2, N2O, O2 and one of the anesthetic agents, halothane, isoflurane, Enflurane, Sevoflurane and Desflurane in any mixture and communicates real time and derived gas information to the Infinity monitors.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K070566, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 31, 2011Z-0281-2012Firm received 2 complaints reporting Ventricular Fibrillation (Vfib) arrhythmias were not recognized by the monitor and did not alarm when the Masimo SET pod is used for SpO2 measurements.Terminated
Jul 29, 2011Z-3047-2011, Z-3048-2011Monitor keys may spontaneously become inoperative or active. This may cause a membrane switch panel malfunction, which could result in the monitor discharging a patient automatically.Terminated
Jun 4, 2010Z-2283-2010, Z-2284-2010Monitor keys may become inoperative or activate spontaneously, causing a membrane switch panel malfunction. This may discharge a patient automatically.Terminated
Nov 9, 2009Z-0582-2010monitor is rebooting/resettingTerminated

Other MHX clearances

The newest 510(k)s for the same product code; their median review was 207 days; see every MHX device

Record

Type
Special
Received
Feb 28, 2007
Review panel
Cardiovascular
Regulation
21 CFR 870.1025

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files