Infinity Delta and Kappa Series Monitors With VF6 Modifications

Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms). 510(k) K060039, from Draeger Medical Systems, Inc. (Danvers, MA). Product code MHX, class 2.

As of Nov 19, 2026, K060039 is an FDA 510(k) held by Draeger Medical Systems, Inc., substantially equivalent on Apr 12, 2006 after 96 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 96 days
  • Recalls: 1
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The ScIO and MultiGas/MultiGas+ modules sample breathing gases from adults and pediatrics. The gas modules continuously measure the content of CO2, N2O, O2 and one of the anesthetic agents, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane in any mixture and communicates real time and derived gas information to the INFINITY monitors.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K060039, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Nov 17, 2006Z-0301-2007Top cover of assembly may separate. The situation can result in the monitor falling, which could result in an injury to the clinician or patient.Terminated

Other MHX clearances

The newest 510(k)s for the same product code; their median review was 207 days; see every MHX device

Record

Type
Special
Received
Jan 6, 2006
Review panel
Cardiovascular
Regulation
21 CFR 870.1025

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.