Hospira Plum A+ Infusion Pump, Model 12391
Pump, Infusion. 510(k) K070398, from Hospira, Inc. (Lake Forest, IL). Product code FRN, class 2.
As of Nov 21, 2026, K070398 is an FDA 510(k) held by Hospira, Inc., substantially equivalent on Apr 24, 2007 after 71 days, with 21 recalls tied to it in FDA's recall database.
- Review time: 71 days
- Recalls: 21
Intended for use in parenteral, enteral, and epidural therapies and the administration of whole blood and blood products The intended use is the same as for the predicate devices.
- Hospira Plum A+ Infusion Pump System V11.5, Hospira Plum A+ 3 Infusion Pump System, V11.5K052052 · Hospira, Inc. · Aug 24, 2005
- Abbott Plum A+ Infusion Pump, Model 12391K024084 · Abbott Laboratories · Dec 31, 2002
| May 7, 2024 | Z-2449-2024, Z-2448-2024Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the… | Open |
| Mar 22, 2023 | Z-1566-2023Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the… | Open |
| Jul 2, 2014 | Z-1072-2015One lot of alarm assemblies used in Plum A+ and Plum A+3 infusion pumps may fail to sound at all volume levels. | Terminated |
| Feb 1, 2013 | Z-2682-2017, Z-2676-2017, Z-2678-2017, Z-2679-2017, Z-2681-2017, Z-2675-2017, Z-2683-2017, Z-2677-2017, Z-2680-2017A possible interruption of therapy when an E321 error code occurs while using the infuser, which is caused by battery issues. | Terminated |
| Feb 1, 2013 | Z-0287-2014Plum A+ infusers have the potential for the distal (occlusion) press sensor pin to break. The distal pressure sensor pin is a part of the overall subsystem that measures the pressure within the distal line of the… | Terminated |
| Dec 7, 2012 | Z-0624-2013The door roller assembly on the Plum A+/A+3 infuser family of infusers has the potential to break which can lead to possible unrestricted flow and/or over-delivery during the removal of the IV administration set's… | Terminated |
| Oct 26, 2012 | Z-0481-2013The volume control knob on some Plum A+ single channel infusers (located on the back of the infuser) may not function as described in the System Operating Manual - the direction for Loud and Quiet may be reversed. The… | Terminated |
| Feb 5, 2012 | Z-0207-2014Hospira has received reports of distal pressure sensor calibration drift on Plum A+ Infusers. The following error codes may be displayed during setup or infusion signifying possible calibration drift: E180/N180,… | Terminated |
| Sep 16, 2011 | Z-3284-2011Hospira has received reports of incorrect seating of the regulator closer on the Plum A+ Infusion Pumps. If the regulator closer does not close the flow regulator actuator when the pump door is opened, and if the… | Terminated |
| Sep 16, 2011 | Z-0009-2012Hospira has received reports of continuous recycling and/or rebooting of Plum A+ infusion pumps when the backlight intensity and/or display contrast settings for the LED display has been adjusted from the original… | Terminated |
| Feb 14, 2011 | Z-1644-2011Hospira has received customer reports of the Plum A+ infusion pumps with no audible alarm conditions, which have been associated with failure of the piezoelectric assembly due to component quality issues. | Terminated |
| Aug 11, 2009 | Z-0024-2010Fire/Shock hazard-- A component used in the device may cause sparks/flashes and poses a fire and shock hazard. | Terminated |
- RemunityPRO+ Pump for Remodulin® (treprostinil) InjectionK253703 · Deka R&D Corp · Jul 24, 2026
- FreedomEdge Infusion System; High-Flo SubQ Needle Set; Precision Flow Rate TubingK252015 · Koru Medical Systems, Inc. · Jan 28, 2026
- Spectrum IQ Infusion System with Dose IQ Safety Software (3570009)K251636 · Baxter Healthcare Corporation · Jul 28, 2025
- SIGMA Spectrum Infusion Pump with Master Drug LibraryK251640 · Baxter Healthcare Corporation · Jul 28, 2025
- MRidium 3870 MRI Infusion Pump System (3870)K242752 · Iradimed Corporation · May 28, 2025
- BD Alaris Infusion System with Guardrails Suite MXK243855 · Carefusion 303, Inc. · Apr 25, 2025
- Plum Solo™ Precision IV PumpK242114 · Icu Medical, Inc. · Apr 2, 2025
- Plum Duo™ Precision IV PumpK242115 · Icu Medical, Inc. · Apr 2, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Feb 12, 2007
- Review panel
- General Hospital
- Regulation
- 21 CFR 880.5725
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files