Giraffe and Panda Warmer
Warmer, Infant Radiant. 510(k) K070377, from Ohmeda Medical (Laurel, MD). Product code FMT, class 2.
As of Nov 19, 2026, K070377 is an FDA 510(k) held by Ohmeda Medical, substantially equivalent on Jul 11, 2007 after 153 days, with 7 recalls tied to it in FDA's recall database.
- Review time: 153 days
- Recalls: 7
- Cited as a predicate by: 3 devices
Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment.
- Infant Warmer SystemK963058 · Ohmeda Medical · Oct 1, 1996
- Ohmeda Medical OmnibedK993407 · Ohmeda Medical · Jan 5, 2000
- Masimo Set RAD-5 Pulse Oximeter, Model RAD-5; Masimo Set RAD-5V Pulse Oximeter, Model RAD-5VK033296 · Masimo Corp. · Feb 19, 2004
- Atom Infa Warmer V-505K002355 · Atom Medical Corp · Oct 10, 2000
- Giraffe and Panda WarmersK123309 · Ohmeda Medical · Feb 5, 2013
- Modification To: Giraffe and Panda WarmerK090697 · Ohmeda Medical · Apr 16, 2009
- Modification To Giraffe and Panda Warmers (Giraffe and Panda Uninterruptible Power Supply)K072157 · Ohmeda Medical · Aug 27, 2007
| Sep 25, 2018 | Z-0483-2019, Z-0482-2019, Z-0484-2019, Z-0481-2019Potential electric leakage affecting the user. | Terminated |
| Nov 18, 2016 | Z-1506-2017Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke,… | Terminated |
| Oct 1, 2013 | Z-1072-2014Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly. | Terminated |
| May 20, 2008 | Z-2227-2008Medical device software may be subject to signal interference of the Hands Free Alarm Silence feature that could impact patient safety. | Terminated |
- Babyroo TN300K243606 · Draeger Medical Systems, Inc. · Jan 17, 2025
- Babyroo TN300K230278 · Draeger Medical Systems, Inc. · May 30, 2023
- Giraffe Omnibed Carestation CS1K213551 · Datex Ohmeda, Inc. · Mar 2, 2022
- Panda iRes Warmer, Giraffe WarmerK201628 · GE Healthcare · Feb 12, 2021
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Feb 8, 2007
- Review panel
- General Hospital
- Regulation
- 21 CFR 880.5130
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files