Warmer, Infant Radiant. 510(k) K070377, from Ohmeda Medical (Laurel, MD). Product code FMT, class 2.

As of Nov 19, 2026, K070377 is an FDA 510(k) held by Ohmeda Medical, substantially equivalent on Jul 11, 2007 after 153 days, with 7 recalls tied to it in FDA's recall database.

  • Review time: 153 days
  • Recalls: 7
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K070377, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Sep 25, 2018Z-0483-2019, Z-0482-2019, Z-0484-2019, Z-0481-2019Potential electric leakage affecting the user.Terminated
Nov 18, 2016Z-1506-2017Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke,…Terminated
Oct 1, 2013Z-1072-2014Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly.Terminated
May 20, 2008Z-2227-2008Medical device software may be subject to signal interference of the Hands Free Alarm Silence feature that could impact patient safety.Terminated

Other FMT clearances

The newest 510(k)s for the same product code; their median review was 118 days; see every FMT device

Record

Type
Traditional
Received
Feb 8, 2007
Review panel
General Hospital
Regulation
21 CFR 880.5130

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files