Arca
Skip to content
Tools
Tools

Abbott Architect Havab-M

Hepatitis A Test (Antibody And Igm Antibody). 510(k) K063329, from Abbott Laboratories (Abbott Park, IL). Product code LOL, class 2.

As of Nov 21, 2026, K063329 is an FDA 510(k) held by Abbott Laboratories, substantially equivalent on Apr 16, 2007 after 164 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 164 days
  • Recalls: 2
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The ARCHITECT HAVAB-M assay is a chemiluminescent microparticle immunoassay (CMIA) for the qualitative detection of IgM antibody to hepatitis A virus (IgM anti-HAV) in human adult and pediatric serum and plasma (dipotassium EDTA, lithium heparin, and sodium heparin) and neonatal serum.

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K063329, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 20, 2011Z-2413-2011The HAVAB-M reagent lot has reduced specificity and may cause an increase in false grayzone and/or false reactive results.Terminated
Dec 30, 2009Z-1596-2010There is a potential to generate a false grayzone or reactive HAVAB-M result when the ARCHITECT AUSAB assay precedes the ARCHITECT HAVAB-M assay.Terminated

Other LOL clearances

The newest 510(k)s for the same product code; their median review was 214 days; see every LOL device

Record

Type
Abbreviated
Received
Nov 3, 2006
Review panel
Microbiology
Regulation
21 CFR 866.3310

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files