Axiom Luminos Drf

Interventional Fluoroscopic X-Ray System. 510(k) K062623, from Siemens Medical Solutions USA, Inc. (Mavern, PA). Product code OWB, class 2.

As of Nov 20, 2026, K062623 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Aug 22, 2007 after 351 days, with 20 recalls tied to it in FDA's recall database.

  • Review time: 351 days
  • Recalls: 20
  • Cited as a predicate by: 7 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The AXIOM Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The AXIOM Iconos dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

7 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K062623, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Dec 21, 2022Z-1166-2023, Z-1164-2023Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default valuesOpen
Apr 13, 2015Z-1668-2015Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing…Terminated
Dec 15, 2014Z-0994-2015Potential for composed images to be flipped before being sent to PACS on systems with software version VE10 and SmartOrtho license. Flipped images may be reversed so the associated annotations, e.g. labels (R/L), may be…Terminated
Sep 2, 2014Z-0003-2015Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause…Terminated
Aug 12, 2014Z-2649-2014It was determined that a potential malfunction may occur on the Luminos dRF system with SW VD10 or on Luminos dRF Max when the table side control is exposed to fluids. Fluid may infiltrate the tableside control panel…Terminated
Jul 7, 2014Z-2292-2014It was discovered that during a RAD examination using Siemens Luminos dRF, Ysio or Uroskop Omnia systems with software version VB10C to VB10F and automatic exposure control, a highly unlikely malfunction can result in…Terminated
Apr 11, 2014Z-1711-2014It was determined that under rare environmental conditions (i.e., extremely dry air in combination with a highly isolated floor) electrostatic discharges (ESD) >8 kV may trigger an unintended movement of the AXIOM…Terminated
Sep 3, 2013Z-0374-2014A potential malfunction and hazard to patients exists when using the Ysio, Luminos Agile, Luminos dRF, and Uroskop Omnia systems with software version VC10, which may result in radiation continuing longer than required…Terminated
Sep 3, 2013Z-0140-2014When using systems operating with software versions VD10A/G during a RAD examination, a malfunction can result in radiation continuing longer than required resulting in an overexposed image that is not of diagnostic…Terminated
Aug 28, 2013Z-1845-2014When using systems operating with software versions VD10A/G during a RAD examination, a malfunction can result in the image appearing very dark, which may prompt unnecessary repetition of the examination.Terminated
May 7, 2013Z-1848-2013Siemens became aware of an unintended behavior when using the Luminos dRF with software version VD10A. The possibility exists that a software error may occur causing the table to tilt into an upright position and may…Terminated
May 6, 2013Z-0016-2014Siemens has discovered through product monitoring a potential malfunction when using system operating with software versions VB10D, VB10F, VB10C and VB10. During a RAD examination the possibility exists that radiation…Terminated
Feb 16, 2012Z-1383-2012During regular product monitoring, Siemens became aware of a potential issue with the AXIOM Luminos dRF. Firm became aware of a potential issue with torque wrenches - during a regular scheduled calibration of the…Terminated
Feb 15, 2012Z-1159-2012Firm has become aware of unintended behavior when using Ysio, AXIOM dRF, AXIOM Luminos Agile/TF and Uroskop Omnia with RAD Fluoro Uro function using software version VC10A. Due to an error in the image pipeline, images…Terminated
Jan 14, 2011Z-1421-2011Firm became aware of a potential malfunction and possible hazard to patients in the event the tilt buttons on the OPTI Grip become stuck under the housingTerminated
Oct 25, 2010Z-0717-2011patient leg can be pinchedTerminated
Jul 13, 2010Z-2222-2010Fluid may enter the multi-functional handle (Opti Grip) when using the Axiom Luminos TF or Sireskop SD system, causing unintended reaction or movement, increased radiation or useless images.Terminated
Sep 19, 2008Z-0326-2009, Z-0327-2009Issue may occur where an intended movement of the joystick may unintentionally initiate an x-ray exposure.Terminated

Other OWB clearances

The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device

Record

Type
Traditional
Received
Sep 5, 2006
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files