Arca
Skip to content
Tools
Tools

Zimmer Trabecular Metal Reverse Shoulder System

Shoulder Prosthesis, Reverse Configuration. 510(k) K052906, from Zimmer, Inc. (Warsaw, IN). Product code PHX, class 2.

As of Nov 18, 2026, K052906 is an FDA 510(k) held by Zimmer, Inc., substantially equivalent on Dec 19, 2005 after 66 days, with 12 recalls tied to it in FDA's recall database.

  • Review time: 66 days
  • Recalls: 12
  • Cited as a predicate by: 11 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Zimmer Trabecular Metal Reverse Shoulder System is indicated for the following

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

11 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K052906, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 22, 2017Z-3144-2017Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly. The anomaly may be on or below the surface and can increase the risk of instrument fracture.Terminated
Oct 17, 2013Z-1079-2014As a result of complaints for disassociation of the TM Reverse Glenosphere from the Baseplate, Zimmer updated the related surgical technique and released a video to provide additional instructions to users on how to…Terminated
Aug 1, 2013Z-2055-2013Zimmer is initiating a lot specific recall of the Trabecular Metal" Reverse Glenosphere Distractor due to the potential of the device not functioning properly should it be utilized without proper lubrication. As a…Terminated
Sep 10, 2010Z-0576-2011, Z-0571-2011, Z-0573-2011, Z-0575-2011, Z-0570-2011, Z-0574-2011The results of an investigation into complaints revealed that Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Humeral implants engaging in the canal earlier than expected could lead to incomplete seating…Terminated
Jul 20, 2006Z-1416-06, Z-1418-06, Z-1417-06Components may not lock together as intended.Terminated

Other PHX clearances

The newest 510(k)s for the same product code; their median review was 132 days; see every PHX device

Record

Type
Traditional
Received
Oct 14, 2005
Review panel
Orthopedic
Regulation
21 CFR 888.3660

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.