Vericis Cardiovascular Image and Information System, Model 4.0
System, Image Processing, Radiological. 510(k) K051649, from Camtronics Medical Systems (Hartland, WI). Product code LLZ, class 2.
As of Nov 18, 2026, K051649 is an FDA 510(k) held by Camtronics Medical Systems, substantially equivalent on Aug 2, 2005 after 42 days, with 17 recalls tied to it in FDA's recall database.
- Review time: 42 days
- Recalls: 17
- Cited as a predicate by: 3 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFVERICIS is a system intended to be used to acquire, store, print, transfer, and archive clinical information from Camtronics and other vendors systems including images, Hemodynamic studies and reports, measurements (via import from DICOM Structured Reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Horizon Medical ImagingK043146 · Mckesson Medical Imaging Company · Jan 4, 2005
- Analytical Review StationK955519 · Camtronics, Ltd. · May 9, 1996
- Echocardiography System Series 95000K992259 · Camtronics, Ltd. · Sep 8, 1999
Cited as a predicate
3 later 510(k) summaries name this device as a predicate- Merge CardioK192276 · Merge Healthcare Incorporated · Sep 20, 2019
- Infinitt Cardiology PacsK082706 · Infinitt Co., Ltd. · Nov 6, 2008
- Syngo Dynamics, Version 5.0K053133 · Siemens Medical Solutions USA, Inc. · Dec 5, 2005
Recalls
Recalls FDA ties to K051649, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Feb 19, 2025 | Z-1401-2025, Z-1400-2025, Z-1398-2025, Z-1399-2025When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may… | Open |
| May 8, 2020 | Z-2403-2020An error in the calculation of the Michigan Reference Ranges can generate an inaccurate Z-Score calculation. The error may reduce the sensitivity of the Z-Score in detecting an abnormality resulting in a false negative. | Terminated |
| Apr 13, 2016 | Z-2123-2017Fetal patient report was automatically pulling prior measurement data for a prior fetus since the fetal study is based on the mother's MRN, resulting in the wrong fetal measurements getting referenced. | Terminated |
| Apr 4, 2016 | Z-1403-2017When taking measurements from images on the Cardio workstation or from the US cart, numbers are not crossing to the report in the correct unit of measure. | Terminated |
| Apr 4, 2016 | Z-1486-2017Cardio study list does not show STAT studies without refreshing. | Terminated |
| Apr 4, 2016 | Z-1415-2017A situation can occur allowing two physicians to access the same study report in EchoIMS when launched from the Cardio Study List without receiving the read-only notification prompt. | Terminated |
| Mar 28, 2016 | Z-0729-2017If images are sent without an order in the system, they will be matched with the latest order on the current patient/modality matching potentially resulting in the matching of the report to the incorrect accession… | Terminated |
| Mar 28, 2016 | Z-1046-2017If a reader selects Multi-study review prior to the original study completing its loading process and then immediately returns to image review of the original study, it is possible that not all current images will be… | Terminated |
| Feb 9, 2016 | Z-2709-2016A report can be confirmed with the incorrect patient demographics resulting in it being saved under the incorrect patient record in the Electronic Medical Record (EMR). | Terminated |
| Jan 30, 2016 | Z-2865-2016Patient information in the header is only found on the first page of the report and not on the subsequent pages. | Terminated |
| Oct 26, 2015 | Z-0555-2016The firm, Merge Healthcare, sent out Merge HEMO V10.0 & Merge CARDIO V10.1 software to 3 customers before the product was completely validated in house. | Terminated |
| Oct 1, 2015 | Z-0730-2017Reporting feature times out after inactivity for more than an hour sending the user back to the study list, which causes all reporting data being entered to be lost. | Terminated |
| Sep 22, 2015 | Z-1496-2017For sites using the Issuer of Patient ID (IPID), the system will display the study list and images from different patients with different IPIDs as though they are for the same patient if they all have the first name,… | Terminated |
| Sep 22, 2015 | Z-1517-2017Users can merge a device import file with an image study that already has a confirmed report, which may result in including information not present at the time of physician interpretation. | Terminated |
Other LLZ clearances
The newest 510(k)s for the same product code; their median review was 145 days; see every LLZ device- 3D Anatomical ModelK262132 · The Jacobs Institute · Nov 5, 2026
- LiLi DICOM Viewer (LiLi )K260454 · Elevate Imaging, Inc. · Nov 4, 2026
- Vitrea Advanced Visualization, Version 7.16K263648 · Canon Medical Informatics, Inc. · Oct 29, 2026
- MIM – KineticIDK261763 · Mim Software, Inc. · Oct 26, 2026
- PV-iCurve ProK261531 · Perception Vision Medical Technology Co., Ltd. · Oct 21, 2026
- Eden PACSK261788 · Higia., Inc. d/b/a Eden · Oct 21, 2026
- Syngo Carbon Space (VA4)K263048 · Siemens Healthineers AG · Sep 29, 2026
- Syneris PACS v.1.0.1.0K261874 · Syneris · Sep 23, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jun 21, 2005
- Review panel
- Radiology
- Regulation
- 21 CFR 892.2050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files