Arca
Skip to content
Tools
Tools

Horizon Medical Imaging

System, Image Processing, Radiological. 510(k) K043146, from Mckesson Medical Imaging Company (Richmond, B.C., CA). Product code LLZ, class 2.

As of Nov 18, 2026, K043146 is an FDA 510(k) held by Mckesson Medical Imaging Company, substantially equivalent on Jan 4, 2005 after 50 days, with 11 recalls tied to it in FDA's recall database.

  • Review time: 50 days
  • Recalls: 11
  • Cited as a predicate by: 5 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Horizon Medical Imaging is a medical image and information management system that is intended to receive, transmit, store, retrieve, display, print and process digital medical images, digital medical video, and associated medical information from various medical imaging systems.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

5 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K043146, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 12, 2016Z-2872-2016The STAT priority icon was not displayed in the Study List when the study was opened.Terminated
Feb 18, 2016Z-1182-2016McKesson has identified a design deficiency where under rare circumstances, imported images/studies may re-use a non-unique image directory. This issue may cause incorrect images to be displayed for a patient.Terminated
May 29, 2014Z-1728-2014Firm received a complaint that an unreported study performed that day could not be opened from the archive.Terminated
May 5, 2014Z-2121-2014The firm has identified a software issue which may cause a discrepancy between the index of images expected by the application, and the actual content of the folder storing the images.Terminated
Apr 17, 2014Z-1617-2014High priority alert icon was not displayed in the Study List when there are no Stat studies present. However, the High priority alert icon is visible when studies marked as Stat priority exist in the Study List.Terminated
Dec 23, 2013Z-0976-2014Images may be archived to an incorrect or invalid archive location resulting in incorrect images or no images being displayed for a study following retrieval from archive.Terminated
Oct 15, 2013Z-0556-2014When merging two patient records into one patient record using McKesson Radiology 12.0, the resulting patient record is missing the Contrast Allergy information for the second patient record.Terminated
Sep 4, 2013Z-2318-2013Software malfunction may occur resulting in a certain combination of server configurations which has the potential to cause image loss.Terminated
May 29, 2012Z-2238-2012There is a potential to save reports in RadReport with an Incorrect Final Status.Terminated
Feb 2, 2011Z-3035-2011Images for one patient may appear in the study of another patient.Terminated
May 7, 2010Z-1735-2010Voice Clip issue: When the user is recording voice clips in Horizon Rad Station, there is a potential for voice clips to be saved to the incorrect study of the same patient.Terminated

Other LLZ clearances

The newest 510(k)s for the same product code; their median review was 145 days; see every LLZ device

Record

Type
Traditional, third-party review
Received
Nov 15, 2004
Review panel
Radiology
Regulation
21 CFR 892.2050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.