Arca
Skip to content
Tools
Tools

Dimension Vista Integrated System

Analyzer, Chemistry (Photometric, Discrete), For Clinical Use. 510(k) K051087, from Dade Behring, Inc. (Newark, DE). Product code JJE, class 1.

As of Nov 18, 2026, K051087 is an FDA 510(k) held by Dade Behring, Inc., substantially equivalent on Jul 11, 2005 after 74 days, with 25 recalls tied to it in FDA's recall database.

  • Review time: 74 days
  • Recalls: 25+
  • Cited as a predicate by: 5 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Dade Behring Dimension Vista™ Integrated system is an in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

5 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K051087, newest first; the 25 newest; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 15, 2021Z-0331-2022, Z-0330-2022Incorrect Default Hemolysis, Icterus or Lipemia (HIL) Index for Five Dimension Vista AssaysTerminated
Dec 12, 2018Z-1014-2019When the Millipore Water Purification Module Progard Pretreatment Packs are used with Dimension Vista Triglyceride assay, patient and QC triglyceride results may be erroneously elevated as a result of contamination.Terminated
Oct 16, 2018Z-0616-2019Siemens Healthcare Diagnostics has identified an issue with Dimension Vista software v.3.7 and v.3.8.1 that may result in aliquot well contamination. If a test is cancelled by the operator while the instrument is…Terminated
Mar 3, 2017Z-1703-2017Siemens Healthcare Diagnostics has confirmed that in isolated cases when ECREA is processed immediately after the weekly automated Acid Clean routine during Probe Test, there is the remote potential for an elevation of…Terminated
Mar 24, 2016Z-1886-2016, Z-1885-2016Discrepant patient results on Dimension Vista Intelligent Lab Systems. Siemens Healthcare confirmed a software defect which, in a very specific set of circumstances, results in the Dimension Vista System omitting an…Terminated
Feb 16, 2016Z-1622-2016Siemens Healthcare Diagnostics has determined that Dimension Vista Blood Urea Nitrogen (BUN) lots 15215AE, 15243BB, 15264BA, 15299BB, 15300BA and 15341AC may exhibit inaccurate patient and/or Quality Control results.…Terminated
May 21, 2015Z-2054-2015Two software issues may occur in Vista software version 3.6.1. Issue #1: Samples may stop processing without notification. Issue #2: Timing issue in Vista 1500 may cause wrong reagent or no reagent delivery. This may…Terminated
Nov 5, 2014Z-0838-2015Customer complaints of discrepant flagged and/or un-flagged patient and QC results. Reagent probe performance may decline more quickly than anticipated and have an effect on assays with similar formulations when…Terminated
Aug 28, 2014Z-0024-2015Siemens Healthcare Diagnostics has determined that Dimension Vista® Mass CKMB Isoenzyme Calibrator (MMB CAL) shows drift during routine stability monitoring that exceeds our acceptance criteria for drift over the shelf…Terminated
Apr 14, 2014Z-1733-2014A small number of Vista Air Compressors were miswired, which if installed and powered up, could cause a fire, or become an electrical shock hazard.Terminated
Oct 7, 2013Z-0851-2014When utilizing the Routine Inventory screen to enter a User Defined Method (EMPTY) Flex the system may assign the User Defined Method Flex to a different Flex that is currently in inventory on the system, and then use…Terminated
Jun 19, 2013Z-2088-2013Siemens is conducting a field correction for Dimension Vista 500 and Dimension Vista 1500 systems using software versions 3.5.1 and 3.6. Under specific rare conditions which require a system reset, there is a potential…Terminated
May 20, 2013Z-0430-2014There is the potential for the vial volume of "in use" Vista calibrators, controls, or sample diluent to incorrectly revert to full volume when removed and reloaded. Dimension Vista(R) 500 and Dimension Vista(R) 1500 -…Terminated
Jan 16, 2012Z-1454-2012Siemens has confirmed customer complaints on Vista(R) software version 3.4 for Below Manufacturer Assay Range error flags associated with QC and patient test results that are within the assay range as defined in the…Terminated
Oct 31, 2011Z-0646-2012, Z-0644-2012, Z-0645-2012Firm has confirmed that specific lots of V-LYTE(R) Fluids may contribute to reduced V-LYTE(R) Integrated Multisensor on-board use life as demonstrated by increases in: Urine sodium QC results within 48 hours after…Terminated
Aug 8, 2011Z-0031-2012Siemens Healthcare Diagnostics has confirmed a low frequency defect in the molding of the Cuvettes that may result in Cuvette Ring and Cuvette Loader jams. It has been determined that the issue is related to a molding…Terminated
Aug 19, 2010Z-0139-2011High potassium content in the affected lots results in the potential for low potassium recovery for Dilution Check, QC and patient samples when run on the Dimension Vista System.Terminated
Jul 23, 2010Z-0121-2011the EasyLink QCFIRST custom rule does not operate as intended.Terminated
May 5, 2010Z-2423-2010Defective batteries used in the Alpha Pinnacle P10 UPS External Battery Pack may overheat and cause a strong sulfuric odor to be released and a corrosive sulfuric acid solution leak to occur.Terminated
Mar 15, 2010Z-2117-2010Vessel pick up errors and vessel jams when the vessel is being loaded. The error halts the instrument and disrupts processing. Vessel flange is thin.Terminated
Nov 24, 2009Z-0980-2010Some vessels may leak. If leakage occurs, patient results will be flagged and are not reportable.Terminated

Record

Type
Traditional
Received
Apr 28, 2005
Review panel
Clinical Chemistry
Regulation
21 CFR 862.2160

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.