Arca
Skip to content
Tools
Tools

GE Datex-Ohmeda Aestiva/5 Mri Anesthesia System

Gas-Machine, Anesthesia. 510(k) K050055, from Datex-Ohmeda, Inc. (Madison, WI). Product code BSZ, class 2.

As of Nov 19, 2026, K050055 is an FDA 510(k) held by Datex-Ohmeda, Inc., substantially equivalent on Feb 4, 2005 after 24 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 24 days
  • Recalls: 2
  • Cited as a predicate by: 4 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Aestiva/5 MRI Anesthesia System provides the functional feature set offered by the conventional Aestiva/5 to the clinician with the added ability to be used in the MR environment. Among those standard Aestiva/5 features is the Datex-Ohmeda user interface, all the ventilation parameters of the SmartVent along with the Aestiva breathing circuit.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K050055, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 19, 2019Z-0114-2020Certain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network. An insufficiently secured terminal server may provide an opportunity for a…Terminated
Dec 15, 2015Z-0679-2016The Drive Gas Check Valve of certain Anesthesia products could become stuck in a fixed open position, which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved, it could…Terminated

Other BSZ clearances

The newest 510(k)s for the same product code; their median review was 265 days; see every BSZ device

Record

Type
Special
Received
Jan 11, 2005
Review panel
Anesthesiology
Regulation
21 CFR 868.5160

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files