Arca
Skip to content
Tools
Tools

Painpump1 and 1 more

Pump, Infusion. 510(k) K042405, from Stryker Corp. (Kalamazoo, MI). Product code FRN, class 2.

As of Nov 20, 2026, K042405 is an FDA 510(k) held by Stryker Corp., substantially equivalent on Oct 14, 2004 after 41 days, with 25 recalls tied to it in FDA's recall database.

  • Review time: 41 days
  • Recalls: 25+
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

PainPump1 delivers controlled amounts of medication directly to the intraoperative site for pain management and/or antibiotic administration. The pump infuses the medication at an hourly flow rate. Medications are infused intraoperatively and postoperatively through intramuscular or subcutaneous routes.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K042405, newest first; the 25 newest; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Sep 2, 2010Z-2593-2010, Z-2594-2010, Z-2591-2010, Z-2592-2010All of the failures lead to the patient receiving less medication than intended. Certain lots of the pain pump have a high failure rate associated with a cracked port housing which leads to internal or external leaks.…Terminated
Aug 16, 2010Z-0096-2011The instructions for use (IFU) has been revised for health care professionals. The indication for use has been changed to remove site specific pain management. The firm has revised the warning to not use the pump around…Terminated
Jun 16, 2009Z-1739-2009, Z-1759-2009, Z-1740-2009, Z-1747-2009, Z-1766-2009, Z-1769-2009, Z-1757-2009, Z-1778-2009, Z-1770-2009, Z-1761-2009, Z-1746-2009, Z-1760-2009, Z-1736-2009, Z-1780-2009, Z-1779-2009, Z-1762-2009, Z-1776-2009, Z-1773-2009, Z-1755-2009, Z-1764-2009The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.Terminated

Other FRN clearances

The newest 510(k)s for the same product code; their median review was 242 days; see every FRN device

Record

Type
Traditional
Received
Sep 3, 2004
Review panel
General Hospital
Regulation
21 CFR 880.5725

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files