Veritas Dry Collagen Matrix, Veritas Dry Or Veritas Psd
Mesh, Surgical. 510(k) K041669, from Synovis Surgical Innovations (St. Paul, MN). Product code FTM, class 2.
As of Nov 18, 2026, K041669 is an FDA 510(k) held by Synovis Surgical Innovations, substantially equivalent on Oct 27, 2004 after 128 days, with 8 recalls tied to it in FDA's recall database.
- Review time: 128 days
- Recalls: 8
- Cited as a predicate by: 3 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFVeritas Dry Staple Line Reinforcement is intended for use as a prosthesis for the surgical repair of soft tissue deficiencies using surgical staplers when staple line reinforcement is needed. Veritas Dry can be used for reinforcement of staple lines during lung and bronchus resections and during bariatric surgical procedures.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Veritas Collagen MatrixK030879 · Synovis Surgical Innovations · Apr 24, 2003
- Peri-Strips Staple Line Reinforcement - Sleeve and Strip Configurations - and Peri-Strips Dry Staple Line ReinforcementK040415 · Synovis Surgical Innovations · Apr 14, 2004
Cited as a predicate
3 later 510(k) summaries name this device as a predicate- Peri-Strips Dry with Veritas Collagen Matrix with Secure Grip Technology (PSDV-SG)K192615 · Synovis Life Technologies, Inc. (Baxter International Inc.) · Dec 17, 2019
- Kerecis SecureMeshK153364 · Kerecis Limited · Aug 19, 2016
- Veritas Collagen Matrix (Dry)K083039 · Synovis Surgical Innovations · Nov 26, 2008
Recalls
Recalls FDA ties to K041669, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Oct 26, 2011 | Z-0274-2012Synovis Surgical Innovations is voluntarily recalling one lot of Peri-Strips Dry with Veritas Collagen Matrix Staple Line Reinforcement (PSD-V) product because some of the devices were mislabeled. | Terminated |
| Mar 16, 2007 | Z-0965-2007, Z-0961-2007, Z-0963-2007, Z-0964-2007, Z-0962-2007Package Integrity/Sterility Compromised-The foil pouch in the Peri-Strips Dry with Veritas Collagen Matrix Circular Staple Line Reinforcement has a potential for minute defects which may compromise the sterility of the… | Terminated |
| Mar 16, 2007 | Z-0960-2007Package Integrity/Sterility Compromised: The foil pouch in the Peri-Strips Dry with Veritas Collagen Matrix Circular Staple Line Reinforcement has a potential for minute defects which may compromise the sterility of the… | Terminated |
| Nov 1, 2005 | Z-0764-06The Peri-Strips Dry with Veritas Collagen Matrix Staple Line Reinforcement devices involved have a small chance that a seal on the outer pouch may not be completely reliable. | Terminated |
Other FTM clearances
The newest 510(k)s for the same product code; their median review was 95 days; see every FTM device- EluPro Antibiotic-Eluting BioEnvelopeK262406 · Boston Scientific · Aug 12, 2026
- Endoform Reconstructive Template - PLGAK250598 · Aroa Biosurgery , Ltd. · Jun 3, 2025
- OviTex PRS (Long-Term Resorbable)K243595 · TELA Bio, Inc. · Dec 19, 2024
- CanGaroo RM Antibacterial EnvelopeK233991 · Elutia, Inc. · Jun 14, 2024
- OviTex PRSK241126 · TELA Bio, Inc. · May 22, 2024
- Peri-Guard and Supple Peri-GuardK223052 · Synovis Life Technologies, Inc. · Apr 7, 2023
- OviTex PRS (Long Term Resorbable)K214070 · Tela Bio · Mar 21, 2023
- VersaWrapK213163 · Alafair Biosciences, Inc. · Oct 29, 2021
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jun 21, 2004
- Review panel
- General, Plastic Surgery
- Regulation
- 21 CFR 878.3300
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files