Peri-Guard and Supple Peri-Guard

Mesh, Surgical. 510(k) K223052, from Synovis Life Technologies, Inc. (St. Paul, MN). Product code FTM, class 2.

As of Nov 17, 2026, K223052 is an FDA 510(k) held by Synovis Life Technologies, Inc., substantially equivalent on Apr 7, 2023 after 190 days, with 4 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Apr 7, 2023
  • Review time: 190 days
  • Recalls: 4

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

PERI-GUARD is intended for use as a prosthesis for the surgical repair of soft tissue deficiencies which include: defects of the abdominal and thoracic wall, gastric banding, muscle flap reinforcement, and hernias (including diaphragmatic, femoral, incisional, inguinal, lumbar, scrotal, and umbilical hernias).

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K223052, newest first; each opens FDA's recall record
Feb 28, 2024Z-1380-2024A new contraindication was added to the IFU which states that Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal…Open, Classified
Feb 28, 2024Z-1379-2024A new contraindication was added to the IFU which states that Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal…Open, Classified
Dec 6, 2023Z-0665-2024Cardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or…Open, Classified
Dec 6, 2023Z-0668-2024Cardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or…Open, Classified

Other FTM Clearances

The newest 510(k)s for the same product code; their median review was 31 days; see every FTM device

Record

Type
Traditional
Received
Sep 29, 2022
Review panel
General, Plastic Surgery
Regulation
21 CFR 878.3300

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files