Pinnacle3 Radiation Therapy Planning System

System, Nuclear Magnetic Resonance Imaging. 510(k) K041577, from Adac Laboratories (Madison, WI). Product code LNH, class 2.

As of Nov 19, 2026, K041577 is an FDA 510(k) held by Adac Laboratories, substantially equivalent on Jun 25, 2004 after 11 days, with 7 recalls tied to it in FDA's recall database.

  • Review time: 11 days
  • Recalls: 7
  • Cited as a predicate by: 4 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Pinnacle 3 Radiation Therapy Planning System is a computer software package intended to provide support for radiation therapy treatment planning for the treatment of benign or malignant disease processes.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K041577, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 5, 2025Z-2570-2025Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet First Prone) orientations.Open
Apr 23, 2015Z-1643-2015A problem (the dose computed in planning mode is calculated incorrectly ) has been detected in the Philips Pinnacle3 software version 10.0 that, if it were to re-occur, could pose a risk for patients or users.…Terminated
Jul 31, 2014Z-2200-2015Philips, Pinnacle Radiation Treatment Planning System version 8.0h, 8.0k, 8.0m, 8.0n, 9 0, 9 2 9 4, 9 6 is being recalled because the dose may be inconsistent with the density of a density-overridden ROl.Terminated
Feb 11, 2014Z-1312-2014Philips Medical Systems have recently determined that a software nonconformance can cause incorrect beam geometry. This issue affects Pinnacle3 Radiation Therapy Planning (RTP) System, software version 9.0 manufactured…Terminated
Oct 5, 2012Z-0206-2013Philips Medical Systems Cleveland), Inc. is recalling Pinnacle3 Virtual Client Connection software version 3.1 due to an improper release and subsequent distribution.Terminated
Nov 18, 2011Z-0594-2012Phillips Medical Systems have recently determined that a software (Pinnacle3 Software version 9.0) nonconformance can cause incorrect beam geometry.Terminated
Aug 21, 2007Z-1212-2007Incorrect Absolute Marking Coordinates when used with PQ-Series scanners due to a software malfunctionTerminated

Other LNH clearances

The newest 510(k)s for the same product code; their median review was 143 days; see every LNH device

Record

Type
Traditional, third-party review
Received
Jun 14, 2004
Review panel
Radiology
Regulation
21 CFR 892.1000

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files