Pinnacle3 Radiation Therapy Planning System
System, Planning, Radiation Therapy Treatment. 510(k) K032724, from Adac Laboratories (Madison, WI). Product code MUJ, class 2.
As of Nov 19, 2026, K032724 is an FDA 510(k) held by Adac Laboratories, substantially equivalent on Oct 9, 2003 after 36 days, with 1 recall tied to it in FDA's recall database.
- Review time: 36 days
- Recalls: 1
- Cited as a predicate by: 4 devices
Pinnacle$^{3}$ Radiation Therapy Planning Software is a computer software package intended to provide support for radiation therapy treatment planning for the treatment of benign or malignant disease processes.
- Adac Pinnacle3TM Radiation Therapy Planning SystemK993923 · Adac Laboratories · Apr 7, 2000
- P3IMRT TMK002237 · Adac Laboratories · Feb 2, 2001
- Crossplan / Accusoft-Xl V.4.04K063482 · Direx Systems Corp. · Dec 14, 2006
- Radionics Xknife RT 3 With Non Stereotactic ModuleK041997 · Radionics, A Division of Tyco Healthcare Group LP · Aug 27, 2004
- Pinnacle3 Radiation Therapy Planning SystemK041577 · Adac Laboratories · Jun 25, 2004
- Standard Imaging E.Imrt CalculatorK040688 · Standard Imaging, Inc. · May 11, 2004
| Started | Recall | Status |
|---|---|---|
| Mar 3, 2004 | Z-0911-04A software defect, relating to the motorized wedge functionality when using an Elekta linear accelerator, can result in incorrect dose delivery to the patient. Also, dose delivery in the wrong location may occur when… | Terminated |
- RT Elements (4.6)K261655 · Brainlab SE · Oct 5, 2026
- Virtual Cone (1.0)K260750 · Varian Medical Systems, Inc. · Jul 8, 2026
- RadOncAIK253050 · Informai · Jun 18, 2026
- Multimodality Simulation Workspace (MM Sim) (v1.0.0)K260207 · Philips Medical Systems Nederland B.V. · May 27, 2026
- eXaSkin PlusK252784 · Anatomical Geometry S.L. · May 22, 2026
- ART.1-USK253078 · Seetreat Pty, Ltd. · May 14, 2026
- syngo.via RT Image Suite VC10K252304 · Varian Medical Systems, Inc. · Mar 18, 2026
- Monaco RTP System (6.3)K252002 · Elekta Solutions AB · Feb 19, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Sep 3, 2003
- Review panel
- Radiology
- Regulation
- 21 CFR 892.5050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files