Sterrad 200 Sterilizer

Sterilizer, Chemical. 510(k) K030429, from Advanced Sterilization Products (Irvin, CA). Product code MLR, class 2.

As of Nov 25, 2026, K030429 is an FDA 510(k) held by Advanced Sterilization Products, substantially equivalent on Aug 18, 2003 after 189 days, with 8 recalls tied to it in FDA's recall database.

  • Review time: 189 days
  • Recalls: 8
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The STERRAD® 200 Sterilizer is designed for sterilization of both metal and nonmetal medical devices at low temperatures. Because the cycle operates within a dry environment and at low temperatures, it is especially suitable for instruments sensitive to heat and moisture.

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K030429, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 4, 2014Z-0844-2015Advanced Sterilization Products is recalling the STERRAD System cassettes because the the leak indicator on the STERRAD System cassettes packaging may not function properly after exposure to hydrogen peroxide.Terminated
Sep 18, 2013Z-0043-2014Advanced Sterilization Products (ASP) is recalling the STERRAD 200 Sterilization System because it may emit an odor or smell into the surrounding environment that is not typical to normal operating conditions.Terminated
Mar 22, 2013Z-1027-2013The recall was initiated because Advanced Sterilization Products (ASP) has determined that the STERRAD 100NX and STERRAD 200 Sterilization Systems may emit an odor or smell into the surrounding environment that is not…Terminated
Mar 13, 2012Z-0998-2013Advanced Sterilization Products (ASP) is recalling the STERRAD 200 System Carriage because it can potentially become dislodged when an excessive load is applied.Terminated
Jan 23, 2012Z-0741-2012ASP determined that an internal filter in the vacuum pump on the STERRAD 200 System is subject to wear over time.Terminated
Feb 25, 2008Z-1630-2008Oil mist filter fails allowing oil mist to be emitted into the vicinity of the sterilizer, which may result in the release of a mist, "haze," or "smoke."Terminated
Nov 12, 2007Z-0845-2008User Guide Update: based on user reports about contact with residual hydrogen peroxide from instruments, pouches and trays after sterilization and subsequent light colored residue on these devices after sterilization…Terminated
Nov 12, 2007Z-0844-2008Outdated Compatibility Lists: Recalling firm has decided to discontinue dissemination of Compatible Medical Device Reference Lists and instrument assessment activitiesTerminated

Other MLR clearances

The newest 510(k)s for the same product code; their median review was 216 days; see every MLR device

Record

Type
Traditional
Received
Feb 10, 2003
Review panel
General Hospital
Regulation
21 CFR 880.6860

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files