Synchron LX I 725 Clinical System
Electrode, Ion Specific, Sodium. 510(k) K023049, from Beckman Coulter, Inc. (Brea, CA). Product code JGS, class 2.
As of Nov 19, 2026, K023049 is an FDA 510(k) held by Beckman Coulter, Inc., substantially equivalent on Oct 4, 2002 after 22 days, with 18 recalls tied to it in FDA's recall database.
- Review time: 22 days
- Recalls: 18
- Cited as a predicate by: 3 devices
The SYNCHRON LX®i 725 System combines the SYNCHRON LX®20 PRO analyzer and the Access® 2 analyzer into a single instrument presentation. Samples are loaded from a single point of entry through a Closed Tube Aliquotter (CTA) unit. The CTA functions as a sample processing manager by aliquotting and routing samples to the Access 2 and/or LX20 PRO analyzer according to programming requirements.
- Modification To Synchron LX 20 Clinical Chemistry SystemK011213 · Beckman Coulter, Inc. · May 16, 2001
- Unicel Dxc 600I Synchron Access Clinical SystemK060256 · Beckman Coulter, Inc. · Feb 27, 2006
- Diazyme Potassium Enzymatic Assay KitK042191 · Diazyme Laboratories · Oct 29, 2004
- Diazyme Sodium Enzymatic AssayK042189 · Diazyme Laboratories · Oct 7, 2004
| Feb 12, 2013 | Z-1034-2013Beckman Coulter is initiating a recall due to a defective newly released aluminum incubator belt pulley that may rub against the light shield, which causes the formation of debris. Debris that is caused by the incubator… | Terminated |
| Nov 1, 2012 | Z-0612-2013It has been determined that, in some cases, pulley drive components have broken parts or have become loose, mispositioned, or disconnected from their drive shafts. When these defects occur, the pulley may not turn at… | Terminated |
| Apr 5, 2011 | Z-2132-2011The recall was initiated because Beckman Coulter has received reports of incorrect electrolyte results on Synchron LX systems due to a variety of maintenance-related hardware issues. | Terminated |
| Oct 22, 2010 | Z-2070-2011This recall was initiated because Beckman Coulter has confirmed that for User-Defined Chemistries (UDRs), the polychrome correction is not automatically updated when wavelengths are modified. | Terminated |
| Oct 5, 2010 | Z-1937-2011The recall was initiated because Beckman Coulter confirmed Access waste bottles that ruptured and expelled liquid waste, exposing the operator to the system liquid waste. Over time, waste bottles can degrade, become… | Terminated |
| Aug 10, 2010 | Z-1619-2011The recall was initiated because Cartridge Glucose (GLU) on UniCel DxC and Synchron LX instruments may give falsely high results for moderately hemolysed samples. The Impact: (1) Neonatal samples drawn by lancets are… | Terminated |
| Apr 15, 2010 | Z-2388-2010UniCel DxC and Synchron LX instruments produced inaccurate (low) GLUCm results. False low results could affect or delay diagnosis or treatment. | Terminated |
| Jan 18, 2010 | Z-2441-2010The recall was initiated due to " A loose Aliquot Probe tube fitting in the Closed Tube Aliquotter (CTA), may result in erroneous Access immunoassay results or Synchron chemistry results". This may result in dilution of… | Terminated |
| Sep 11, 2008 | Z-1374-2012The recall was initiated because Beckman confirmed a customer report that a timing conflict can occur while aspirating samples for the Access Total BhCG assay. | Terminated |
| Apr 3, 2008 | Z-2385-2008The field action was initiated after the firm confirmed reports of premature failure of the waste pump tubing in the Access and Access 2 Immunoassay Systems. Premature failure of the waste pump tubing may negatively… | Terminated |
| Oct 17, 2007 | Z-1183-2008Electrical grounding failure: During manufacturing, internal testing found one instrument wlth an intermittent failure of the electrical grounding. The source of the problem was traced to inadequate soldering. | Terminated |
| Apr 5, 2007 | Z-1096-2008Intermittent Failure of Stirrer Motor-Beckman Coulter has determined that incorrect Glucose results on LX20 or DxC instruments have been related to an intermittent failure of the stirrer motor. -Stirrer motors can stall… | Terminated |
| Mar 15, 2006 | Z-0901-06It was confirmed that the ORDAC (Over Range Detection And Correction) feature (auto & manual) does not perform correctly for the glucose cup chemistry (GLUCm) when sample type of 'Other' is selected w/operating software… | Terminated |
| Sep 6, 2005 | Z-0009-06Synchron LX Systems Operating Software Version 4.5/v4.6 CBD (Chemistry Database) introduced an adjustment to the urine Uric Acid database calculation scheme which for some patients, depending on the volume of 24 hour… | Terminated |
| Oct 10, 2003 | Z-1098-04Numeric results are erroneously accepted by the in-house information system without qualifying symbols "<" or ">". In addition, the in-house system may interpret the results incorrectly. | Terminated |
| Oct 7, 2003 | Z-1116-04'Initial Rate High' parameter for User Defined Reagents does not perform as indicated in the Synchron LX Clinical Systems Operations Manual. | Terminated |
| Sep 18, 2003 | Z-0096-04Sample wheel may home to the wrong position. | Terminated |
| Jan 30, 2003 | Z-0693-03False results from secondary pour off tubes | Terminated |
- ABL90 FLEX PLUS SystemK252475 · Radiometer Medicals Aps · May 1, 2026
- i-STAT CG8+ cartridge with the i-STAT 1 SystemK230275 · Abbott Point of Care, Inc. · Oct 27, 2023
- A-LYTE® Integrated Multisensor (IMT Na K Cl)K222438 · Siemens Healthcare Diagnostics, Inc. · Jul 25, 2023
- ISE Reagents, Glucose, CRP Latex, DxC 500 AU Clinical Chemistry AnalyzerK220977 · Beckman Coulter Laboratory Systems (Suzhou) Co., Ltd. · Jul 20, 2023
- GLP systems TrackK213486 · Abbott Laboratories · Mar 10, 2022
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Sep 12, 2002
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1665
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files