Arca
Skip to content
Tools
Tools

Restoration Modular System

Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented. 510(k) K022549, from Howmedica Osteonics Corp. (Allendale, NJ). Product code LZO, class 2.

As of Nov 19, 2026, K022549 is an FDA 510(k) held by Howmedica Osteonics Corp., substantially equivalent on Sep 23, 2002 after 53 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 53 days
  • Recalls: 3
  • Cited as a predicate by: 7 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Restoration™ Modular System is intended to be used for primary or revision total hip arthroplasty, as well as in the presence of severe proximal bone loss. The stems are intended to be used with Howmedica Osteonics femoral heads, unipolar and bipolar components, and acetabular components. These femoral stems are designed to be press fit into the proximal femur.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

7 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K022549, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 2, 2012Z-2487-2012Stryker Orthopaedics has become aware of a burr, or a metal shaving, observed in the internal threaded portion of the Cone Body Trial. The typical size of the shaving observed was 1-2 mm in length and 0.037 mm squared…Terminated
Apr 2, 2009Z-1642-2009Packaging burst test values for the Restoration Modular Cone Body family were found to be incorrectly transferred from the finalized validation to the work specification. Testing showed that burst test results for…Terminated
Nov 21, 2008Z-0884-2009Stryker became aware of a lot for lot mix up.Terminated

Other LZO clearances

The newest 510(k)s for the same product code; their median review was 102 days; see every LZO device

Record

Type
Traditional
Received
Aug 1, 2002
Review panel
Orthopedic
Regulation
21 CFR 888.3353

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files