Arca
Skip to content
Tools
Tools

Baxter Accura System, Model 5M5660

Dialyzer, High Permeability With Or Without Sealed Dialysate System. 510(k) K021615, from Baxter Healthcare Corp (Mcgaw Park, IL). Product code KDI, class 2.

As of Nov 20, 2026, K021615 is an FDA 510(k) held by Baxter Healthcare Corp, substantially equivalent on Nov 18, 2002 after 186 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 186 days
  • Recalls: 3
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Baxter Accura System is indicated for continuous solute and/or fluid removal in patients with acute renal failure or fluid overload. The Baxter Accura System may also be used in Therapeutic Plasma Exchange (TPE) therapies.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K021615, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 29, 2005Z-0153-06A software anomaly that occurs when the Reset Totals option is selected during CVVHD therapy with a diaylsate flow rate of less than 600 mL/hour, or CVVH therapy with a predilution flow rate of less than 600 mL/hour,…Terminated
Jul 13, 2005Z-1380-05In the event therapy is terminated, as part of routine instrument programming or therapy maintenance, the Accura instrument may experience difficulty in performing a successful self-test (failure) upon an immediate…Terminated
Jul 26, 2004Z-1338-04The Accura machine display screen may go blank during patient treatment. There is a potential risk to the patient in that the operator may try to manipulate the controls in response to a blank screen, which could affect…Terminated

Other KDI clearances

The newest 510(k)s for the same product code; their median review was 189 days; see every KDI device

Record

Type
Traditional
Received
May 16, 2002
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.5860

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files