Easystat PH, PCO2, PO2, Hct, Na+, K+, Ca++ Analyzer
Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph. 510(k) K021515, from Medica Corp. (Bedford, MA). Product code CHL, class 2.
As of Nov 19, 2026, K021515 is an FDA 510(k) held by Medica Corp., substantially equivalent on Jul 16, 2002 after 67 days, with 1 recall tied to it in FDA's recall database.
- Review time: 67 days
- Recalls: 1
- Cited as a predicate by: 1 device
The EasyStat analyzer is designed for clinical laboratory use, making direct measurements of pH (hydrogen ion activity), PCO₂ (partial pressure of carbon dioxide), PO₂ (partial pressure of oxygen), Hct (Hematocrit), Na⁺ (sodium), K⁺ (potassium), and Ca²⁺ (ionized calcium) on whole blood samples from syringes or capillary tubes.
- EasyStat 300K211559 · Medica Corporation · Jan 27, 2023
| Started | Recall | Status |
|---|---|---|
| Dec 31, 2025 | Z-1370-2026A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes. | Open |
- ABL90 FLEX PLUS System, safeCLINITUBESK252207 · Radiometer Medicals Aps · Apr 10, 2026
- i-STAT CG4+ cartridge with the i-STAT 1 SystemK244014 · Abbott Point of Care, Inc. · May 2, 2025
- ABL90 FLEX PLUS SystemK240998 · Radiometer Medicals Aps · Dec 13, 2024
- i-STAT CG8+ cartridge with the i-STAT 1 SystemK230285 · Abbott Point of Care, Inc. · Oct 27, 2023
- Stat Profile Prime Plus Analyzer SystemK221900 · Nova Biomedical Corporation · Sep 29, 2023
- i-STAT G3+ cartridge with the i-STAT 1 SystemK223857 · Abbott Point of Care, Inc. · Sep 15, 2023
- GEM Premier 7000 with IQM3K223608 · Instrumentation Laboratory Company · Aug 10, 2023
- EasyStat 300K211559 · Medica Corporation · Jan 27, 2023
- Type
- Traditional
- Received
- May 10, 2002
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1120
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files