Philips M5066A and M5068A

Automated External Defibrillators (Non-Wearable). 510(k) K020715, from Philips Medical Systems (Seattle, WA). Product code MKJ, class 3.

As of Nov 20, 2026, K020715 is an FDA 510(k) held by Philips Medical Systems, substantially equivalent - subject to tracking reg. on Nov 8, 2002 after 248 days, with 12 recalls tied to it in FDA's recall database.

  • Review time: 248 days
  • Recalls: 12
  • Cited as a predicate by: 6 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The M5066A is designed to be used on a person in sudden cardiac arrest, who is: - unresponsive when shaken, and - not breathing normally. If in doubt, apply the pads. If the victim is an infant or child younger than eight years old or weighs less than 55 lbs (25 kg), you should use the special infant/child pads. If the child appears older/larger, use the adult pads.

Cited as a predicate

6 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K020715, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Sep 10, 2019Z-1271-2020, Z-1272-2020Affected devices contain a Printed Circuit Assembly (PCA) that was not manufactured/repaired in accordance to quality standards.Terminated
Aug 8, 2018Z-2549-2019The recall is being initiated due to incorrect age range labeling (8+ yrs) on the foil packaging for infant/child pad cartridges. The infant/child pad cartridge is designed for use with the defibrillator An Infant/Child…Terminated
Mar 14, 2018Z-1914-2018An issue with one of the electric components (a resistor) could result in the device not delivering a shock when needed.Terminated
Feb 6, 2014Z-1232-2014Philips is recalling HeartStart Home and OnSite (HS1) automated external defibrillators (AEDs) because one or both contact pins may be contaminated by residue from the soldering process. This contamination could prevent…Terminated
Jul 26, 2012Z-0643-2013Philips determined that it is important to clarify information in the Owner's Manual and keep customers informed about the maintenance of their FRx (models 861304 and 861305), OnSite (models M5066A and M5067A), or…Terminated
Apr 30, 2010Z-1916-2010, Z-1917-2010Seventeen AEDs failed production line testing (Final Acceptance Test).Terminated
Nov 2, 2009Z-1194-2010, Z-1195-2010Potential for failure of a capacitor during use which would prevent the AED from delivering effective defibrillation therapy when indicated.Terminated
Apr 30, 2008Z-1780-2008, Z-1779-2008Buttons Sticking - The On/Off and Shock buttons on the front panel of the device may stick in place when pressed and fail to respond to the button press, causing a delay in the delivery of a defibrillation shock when…Terminated

Record

Type
Abbreviated
Received
Mar 5, 2002
Review panel
Cardiovascular
Regulation
21 CFR 870.5310

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files