Arca
Skip to content
Tools
Tools

Roche Cobas Taqman Analyzer

Analyzer, Chemistry, Micro, For Clinical Use. 510(k) K012966, from Roche Molecular Systems, Inc. (Pleasanton, CA). Product code JJF, class 1.

As of Nov 20, 2026, K012966 is an FDA 510(k) held by Roche Molecular Systems, Inc., substantially equivalent on Feb 1, 2002 after 150 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 150 days
  • Recalls: 3
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The COBAS TaqMan Analyzer is a fully automated amplification and detection system for nucleic acids using 5' nuclease technology. The COBAS TaqMan Analyzer is intended to be used by laboratory professionals trained in laboratory techniques and on the use of the Analyzer.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K012966, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 24, 2004Z-1119-04, Z-1120-04Improper seating of a fiber optic cable which may cause pixel cross-talk leading to falsely elevated test results or cases of misalignment to sample-to-result mismatch.Terminated
Mar 6, 2003Z-0917-03A false positive result may be reported.Terminated

Record

Type
Traditional
Received
Sep 4, 2001
Review panel
Clinical Chemistry
Regulation
21 CFR 862.2170

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files