Modification To Olympus AU2700 Clinical Chemistry Analyzer
Analyzer, Chemistry (Photometric, Discrete), For Clinical Use. 510(k) K003721, from Olympus America, Inc. (Irving, TX). Product code JJE, class 1.
As of Nov 21, 2026, K003721 is an FDA 510(k) held by Olympus America, Inc., substantially equivalent on Dec 18, 2000 after 14 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 14 days
- Recalls: 4
- Cited as a predicate by: 2 devices
- ISE Reagent, Glucose, CRP Latex, DxC 700 AU Clinical Chemistry AnalyzerK161837 · Beckman Coulter, Inc. · Dec 16, 2016
- AU5800(R) Chemistry AnalyzerK112412 · Beckman Coulter, Inc. · Dec 23, 2011
| Sep 9, 2013 | Z-0152-2014The recall was initiated because Beckman Coulter has confirmed that some AU Clinical Chemistry Analyzer ISE units can experience premature ISE REF (Reference) valve failure. Premature valve failure can be detected by… | Terminated |
| Apr 26, 2013 | Z-1430-2013Beckman Coulter is recalling AU2700/AU2700 Plus & AU5400 Clinical Chemistry Analyzers becuase it was determined that the tubing/joint connections associated with these MODs may leak concentrated detergent. Beckman… | Terminated |
| Jul 12, 2010 | Z-0976-2012, Z-0974-2012The recall was initiated because Beckman has confirmed that the AU Clinical Chemistry Analyzers may experience cuvette overflow. | Terminated |
- Type
- Special
- Received
- Dec 4, 2000
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.2160
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files