• Further details on additional microbiological plates that were not initially enumerated and the results that your firm ultimately obtained for these plates. Also, summarize all lots made without sufficient environmental monitoring on the (b)(4) and (b)(4). Provide risk assessments for any potentially affected products marketed to the United States.
• Your investigations of multiple deviations from action limits for ISO 5 and other clean areas.
• A thorough, independent assessment with corrective actions and preventive actions (CAPA) for your environmental monitoring and personnel monitoring programs. For instance, your remediation should include adequate sampling procedures, media suitability, sample accountability (e.g., identification, storage, logging, analysis dates/times), appropriate locations and frequencies, proper responses to alert and action limits, routine identification of microbes isolated from cleanroom and personnel samples, and various other elements of an effective program.
• A comprehensive, independent review of your laboratory practices, methods, equipment, and analyst competencies. Based on this review, provide a detailed CAPA plan to fully remediate your laboratory system. Your plan should also include the process you will use to evaluate the effectiveness of the implemented CAPA.
• A comprehensive identification of all contamination hazards in your aseptic processes, equipment, and facilities. Provide an independent risk assessment that covers, among other things, all human interactions with the ISO 5 area, equipment placement and ergonomics, air quality in the ISO 5 area and surrounding room, facility layout, personnel flow, and material flow.
• A retrospective review into all missing or delayed stability testing that is intended to support the shelf-life of each of your U.S. products.
• An impact assessment for the missed or delayed stability testing, and an updated summary of all stability data (i.e., data obtained for each testing) supporting each of your U.S. products.
• A comprehensive assessment and CAPA to ensure the adequacy of your stability program. Your CAPA should include, but should not be limited to, correcting the root causes of the missed and delayed stations, ensuring adequate number of qualified quality control personnel, improved procedures, and comprehensive personnel training.
• A comprehensive investigation into the extent of the inaccuracies in data records and reporting, including results of the data review for drugs distributed to the United States. Include a detailed description of the scope and root causes of your data integrity lapses.
• A current risk assessment of the potential effects of the observed failures on the quality of your drugs. Your assessment should include analyses of the risks to patients caused by the release of drugs affected by a lapse of data integrity, and risks posed by ongoing operations.
• A management strategy for your firm that includes the details of your global corrective action and preventive action plan. The detailed corrective action plan should describe how you intend to ensure the reliability and completeness of all data generated by your firm, including microbiological and analytical data, manufacturing records, and all data submitted to FDA.