Arca
Skip to content
Tools
Tools

Warning Letter to Longood Medicine (Beijing) Co Ltd

Unspecified. See Longood Medicine (Beijing)'s company page or this site's inspections.

On Aug 27, 2018, FDA's CDER sent Longood Medicine (Beijing) Co Ltd a warning letter citing sterility assurance, process validation, quality unit.

  • Letter date: Aug 27, 2018
  • Issued by: CDER

The Letter

As posted on FDA.gov ↗; cited 21 CFR sections link to their page.

• Your qualification procedures for assessing the suitability and competence of potential contractors before outsourcing, and for ongoing monitoring and review of the performance of the contract facility to identifying and implement any needed improvements;

• A detailed plan for ensuring that the process your contractor uses to terminally sterilize all products, including Chlora-Cleanze Proprep applicators, is adequately validated;

• Complete sterilization records from your contractor for Chlora-Cleanze Proprep applicators, lot 20170601. If you do not have sufficient evidence that the lot in question or any lot intended for the U.S. market was adequately sterilized, indicate the corrective actions you will take, such as customer notifications and product recalls.

• A corrective action and preventive action (CAPA) plan to fully remediate your review and release procedures for finished drug products. Your procedures should require that third-party contractors provide sufficient production and control records to demonstrate compliance with established, approved written procedures and specifications before a lot is released or distributed by your quality unit.

• A comprehensive, independent assessment of documentation systems used throughout your manufacturing and laboratory operations to determine where documentation practices are deficient. Include a detailed CAPA plan that systemically remediates deficient documentation practices, and ensures you retain complete and accurate records.

• A comprehensive CAPA for your training program to ensure all staff at your facility are fully trained in CGMP. Provide a fully remediated training program for your entire operation. Place special emphasis on assuring that all staff involved in any CGMP function are trained and competent in acceptable recordkeeping practices, such as retaining all records, completing records contemporaneously, documenting any error in records, and ensuring that all procedures are approved by quality unit.

• “Chlorhexidine gluconate 2% w/v…Antiseptic”

• “Isopropyl Alcohol 70% w/v…Antiseptic”

• “Antiseptic Solution”

• “Use prior to surgery to prepare patient’s skin. Helps in the reduction of potentially harmful bacteria”

Recent Letters Citing Sterility assurance

See them all

Record

Recipient
Mr. Shenglin Gao
FDA site ID (FEI)
3011090864