Warning Letter to Fresenius Kabi Oncology Limited (Baddi)

Drugs. See Fresenius Kabi Oncology Limited (Baddi)'s company page or this site's inspections.

On Dec 17, 2017, FDA's CDER sent Fresenius Kabi Oncology Limited (Baddi) a warning letter citing aseptic processing, sterility assurance, out-of-specification.

  • Letter date: Dec 17, 2017
  • Issued by: CDER

The Letter

As posted on FDA.gov ↗; cited 21 CFR sections link to their page.

• The sterility test was performed using a (b)(4) filtration system. This (b)(4) testing system was used inside an ISO-5 closed restricted access barrier system (cRABS). Both provisions significantly minimize the potential for introduction of adventious contamination during a sterility test.

• No microbiological contamination was observed in the negative controls.

• No aseptic breaches were observed during the sterility test.

• Environmental monitoring data did not indicate that the sterility testing cRABS had a loss of control.

• Other materials used in performing the sterility test were also subjected to additional sporicidal disinfections.

• Your plans and procedures to ensure that future sterility failure investigations include

• An assessment of your overall system for investigations of deviations, atypical events, complaints, out-of-specification results, and failures. Your corrective action and preventive action (CAPA) plan should include, but not be limited to, improved rigor in reviewing the sources of variation in your operation that may cause deviations, failures, or defects.

• A detailed explanation of how (b)(4) sterility samples and other sterility testing materials are immersed in a sporicidal disinfectant, whether these materials are completely immersed, and a CAPA if the latter is done.

• A review of your aseptic processing operation. Provide a formal assessment of microbiological contamination risks in your current process, equipment, and facility, and a CAPA plan to address identified hazards.

The Firm's Other Letters

Same company or the same FDA site, newest first

Recent Letters Citing Aseptic processing

See them all

Record

Recipient
Mats Henriksson
FDA site ID (FEI)
3006210232

Tags are assigned by Arca from the letter's text. A closeout letter means FDA found the firm's corrections adequate; no closeout does not mean the issues are unresolved. FDA warning letters

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.