• Per the report’s Section 1.1, the power input test was conducted at an exposure time of (b)(4); however, no justification was documented to demonstrate this time as the worst-case scenario for highest power.
• Per the report’s Section 3.1, exposure time accuracy was tested at exposure (b)(4). Per your firm’s CEO, Raymond Ryu, the test standard (which was not referenced in the report) requires exposure time test the device at the lowest exposure time. However, the test was also (b)(4) was documented conducting the (b)(4), the maximum exposure time for the device.
• The report does not reference the standards followed for the testing.
• Your firm has not maintained results for the testing conducted in support of the design change.
• The report does not state the equipment that was used for testing.
• (b)(4), dated 2016-05-25, which states “X-RAY is not generated”
• (b)(4), dated 2016-07-11, which states “X-RAY is not exposed”