FDA establishment 3030254128, Englewood Cliffs, New Jersey, United States. Part of Reset Technology's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
Inspections: 1
Official action indicated: 1
Warning letters: 1
Inspections by fiscal year
FY2018 to FY2027 to date; 0 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 801.20(a). Label to bear a unique device identifier The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
21 CFR 830.300(a). Devices subject to device identification GUDID data submission requirements. The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.
21 CFR 820.70(g). Equipment design and installation Equipment used in the manufacturing process has not been appropriately placed to facilitate maintenance, adjustment, cleaning, and use.
21 CFR 820.70(i). Software validation for automated processes Software used as part of the quality system has not been validated for its intended use according to an established protocol.
21 CFR 820.30(a). Design control - no procedures Procedures for design control have not been established.
21 CFR 820.90(b)(2). Product rework procedures, Lack of or inadequate procedures Procedures for rework of nonconforming product have not been established.
21 CFR 820.90(a). Nonconforming product, Lack of or inadequate procedures Procedures have not been adequately established to control product that does not conform to specified requirements.
21 CFR 820.25(b). Training - Lack of or inadequate procedures Procedures for training and identifying training needs have not been adequately established.
21 CFR 820.70(c). Environmental control Lack of or inadequate procedures Procedures to control environmental conditions have not been established.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 820.75(a). Documentation Process validation activities have not been documented.
Official Action Indicated (OAI)
Warning letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports