FDA establishment 3027535342, Bellevue, Washington, United States. Part of SeniorLife Technologies's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
Inspections: 1
Official action indicated: 1
Form 483s posted: 1
Warning letters: 1
Inspections by fiscal year
FY2018 to FY2027 to date; 1 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 820.25(b). Training - Lack of or inadequate procedures Procedures for training and identifying training needs have not been established.
21 CFR 820.30(a). Design control - no procedures Procedures for design control have not been established.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been established.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
21 CFR 820.20(b)(3). Management representative No management representative had been appointed to ensure that quality system requirements are met, and to report to management on the performance of the quality system.
21 CFR 803.17. Lack of Written MDR Procedures Written MDR procedures have not been developed.
21 CFR 820.22. Quality audits - Lack of or inadequate procedures Procedures for quality audits have not been established.
Official Action Indicated (OAI)
Form 483 observations
What FDA investigators wrote up at the end of an inspection, from the 483s FDA posted; newest first
A system to assign unique device identification (UDI) to the medical device was not documented.
Design and development procedures were not documented.
Procedures for validation of processes did not adequately include requirements for records and use of specific methods, procedures, and acceptance criteria.
Outsourced processes that affect product conformity to requirements were not monitored and controlled.
Internal audits have not been adequately conducted at planned intervals to determine whether the QMS is effectively implemented and maintained.
Quality management system processes were not adequately managed.
Documents required by the quality management system have not been controlled.
Warning letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports