FDA establishment 3015767533, San Juan, Puerto Rico, United States. Every inspection FDA has posted for the site, with its outcome and cited regulations.
Inspections: 2
Official action indicated: 1
Warning letters: 0
Inspections by fiscal year
FY2018 to FY2027 to date; 1 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
Biologics · Blood and Blood ProductsBiologics · Human Cellular, Tissue, and Gene Therapies
3 cited references
21 CFR 1271.230(b). Reduced risk representation Written representation that processing methods reduce the risk of transmission of communicable disease by HCT/Ps was not based on a fully validated or verified process.
21 CFR 1271.160(b)(6). Deviations--evaluation, cause, corrective action Investigation of deviations related to core CGTP requirements did not include corrective action(s) to address the deviation and prevent recurrence.
21 CFR 1271.180(a). Procedures to meet core CTGP Procedures appropriate to meet core CGTP requirements for all steps that you perform in the manufacture of HCT/Ps were not documented.
Voluntary Action Indicated (VAI)
Sep 18, 2019
Biologics · Human Cellular, Tissue, and Gene Therapies
6 cited references
21 CFR 1271.195(c). Monitoring-microorganisms where appropriate Environmental conditions were not monitored for microorganisms.
21 CFR 1271.195(a). Adequate control, proper conditions Environmental conditions existed in which contamination or cross contamination of HCT/Ps or equipment could occur, and environmental conditions were not adequately controlled and proper conditions for operations were not provided.
21 CFR 1271.230(a). Validation & approval--established procedures Processes with results which could not be fully verified by inspection and tests, were not validated and approved according to established procedures.
21 CFR 1271.270(a). Records incomplete Records were not detailed as necessary to provide a complete history of work performed.
21 CFR 1271.160(b)(6). Deviations--Investigation, documenting, trending The quality program does not include the investigation of HCT/P deviations relating to core CTGP requirements.
21 CFR 1271.190(a). Size, construction, location--prevent mix-ups etc. Facilities were not of suitable size and construction to prevent contamination of HCT/Ps with communicable disease agents and ensure the orderly handling of HCT/Ps without mix-ups.
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports