FDA establishment 3015730775, Lawrenceville, New Jersey, United States. Part of LaseTech's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
Inspections: 2
Official action indicated: 1
Warning letters: 2
Inspections by fiscal year
FY2018 to FY2027 to date; 1 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
Devices · Radiation Control and Health Safety ActDevices · Postmarket Assurance: Devices; Devices · Compliance: Devices
10 cited references
21 CFR 830.300(a). Devices subject to device identification GUDID data submission requirements. The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.
21 CFR 801.20(a). Label to bear a unique device identifier The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.
21 CFR 820.30(g). Design validation- Lack of or inadequate procedures Procedures for design validation have not been adequately established.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
21 CFR 820.181. DMR - not or inadequately maintained A device master record has not been adequately maintained.
21 CFR 820.40. Procedures not adequately established or maintained Document control procedures have not been adequately established.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 820.25(b). Training - Lack of or inadequate procedures Procedures for training and identifying training needs have not been adequately established.
21 CFR 1010.3(a)(2)(ii). Manufacture date abbreviated or unapproved date coding The date of manufacture provided on the identification label is abbreviated.
21 CFR 820.184. DHR content The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
Official Action Indicated (OAI)
Feb 10, 2022
Devices · Compliance: DevicesDevices · Radiation Control and Health Safety Act
3 cited references
21 CFR 820.25(b). Training - Lack of or inadequate procedures Procedures for training and identifying training needs have not been established.
21 CFR 820.184. DHR content The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
21 CFR 820.30(f). Design verification - documentation The design verification results, including method(s) and the date, were not adequately documented in the design history file.
Voluntary Action Indicated (VAI)
Warning letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports