FDA establishment 3015283772, Grassobbio, Italy. Part of Mectronic Medicale's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
Inspections: 1
Official action indicated: 1
Warning letters: 1
Inspections by fiscal year
FY2018 to FY2027 to date; 0 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
Devices · Radiation Control and Health Safety ActDevices · Compliance: Devices; Devices · Postmarket Assurance: Devices
11 cited references
21 CFR 820.75(a). Lack of or inadequate process validation A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
21 CFR 820.200(a). Servicing - Lack of or inadequate procedures Procedures or instructions for performing servicing activities and verifying that servicing meets specified requirements have not been adequately established.
21 CFR 1040.10(g)(3). Class IV "Danger" label Each Class IV laser product does not have affixed a label bearing the DANGER logotype specified in the regulation.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
21 CFR 820.30(a). Design control - no procedures Procedures for design control have not been established.
21 CFR 803.17. Lack of Written MDR Procedures Written MDR procedures have not been developed.
21 CFR 820.184. DHR content The device history record does not demonstrate that the device was manufactured in accordance with 21 CFR 820.
21 CFR 820.80(a). Lack of or inadequate procedures - Acceptance activities Procedures for acceptance activities have not been adequately established.
21 CFR 820.90(a). Nonconforming product, Lack of or inadequate procedures Procedures have not been adequately established to control product that does not conform to specified requirements.
21 CFR 820.25(b). Training - Lack of or inadequate procedures Procedures for training and identifying training needs have not been adequately established.
21 CFR 820.22. Quality audits - auditor independence Individuals who conduct quality audits have direct responsibility for the matters being audited.
Official Action Indicated (OAI)
Warning letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports