FDA establishment 3014615028, Miaoli County, Taiwan. Part of Pacific Hospital Supply's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
21 CFR 820.70(a). Process control procedures, Lack of or inadequate procedures Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
21 CFR 820.90(a). Nonconforming product, Lack of or inadequate procedures Procedures have not been established to control product that does not conform to specified requirements.
21 CFR 820.75(a). Lack of or inadequate process validation A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
Official Action Indicated (OAI)
Aug 29, 2013
Devices · Compliance: Devices
4 cited references
21 CFR 820.40. Lack of procedures, or not maintained Document control procedures have not been established.
21 CFR 820.70(g). Equipment Installation, Placement, Specified Requirements The appropriate design, construction, placement, and installation of manufacturing equipment have not been ensured.
21 CFR 820.75(b)(2). Documentation of validated process performance There is inadequate documentation of monitoring and control methods and data and the major equipment used for a validated process.
21 CFR 820.140. Lack of or inadequate procedures for handling Procedures for product handling have not been established.
Voluntary Action Indicated (VAI)
Warning letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports