Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.
FDA establishment 3014334038, Mansfield, Massachusetts, United States. Part of Integra LifeSciences's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
Inspections: 1
Official action indicated: 1
Last inspection ended: Jun 5, 2024
Warning letters: 1
Import refusals 2014 to 2023: 1, first Jul 27, 2020
Inspections by Fiscal Year
FY2018 to FY2027 to date; 1 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 820.70(c). Environmental control Lack of or inadequate procedures Procedures to control environmental conditions have not been adequately established.
21 CFR 820.90(a). Nonconforming product control Products that do not conform to specifications are not adequately controlled.
21 CFR 820.100(b). Documentation Corrective and preventive action activities and/or results have not been documented.
21 CFR 820.150. Lack of or inadequate procedures for storage. Procedures for the control of storage areas and stock rooms have not been adequately established.
21 CFR 820.70(e). Contamination control, Lack of or inadequate procedures Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been adequately established.
21 CFR 820.75(a). Lack of or inadequate process validation A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
21 CFR 820.80(d). Lack of or inadequate final acceptance procedures Procedures for finished device acceptance have not been adequately established.
Official Action Indicated (OAI)
Warning Letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm.FDA Data Dashboard, warning letters and enforcement reports