FDA establishment 3011053282, Nesconset, New York, United States. Part of LifeVac's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
Inspections: 2
Official action indicated: 0
Warning letters: 2
Inspections by fiscal year
FY2018 to FY2027 to date; 0 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 820.80(b). Lack of or inadequate receiving acceptance procedures Procedures for acceptance of incoming product have not been established.
21 CFR 820.30(g). Design validation acceptance criteria Acceptance criteria were not established prior to the performance of validation activities.
21 CFR 820.30(d). Design output - review and approval Design output was not reviewed and approved before release.
21 CFR 820.30(e). Design review - Lack of or inadequate procedures Procedures for design review have not been established.
21 CFR 820.30(c). Design input - documentation Design input requirements were not adequately documented.
21 CFR 820.30(c). Design input - Lack of or inadequate procedures Procedures for design input have not been established.
21 CFR 820.30(b). Design plans - Lack of or inadequate Design plans that describe or reference the design and development activities and define responsibility for implementation have not been established.
21 CFR 820.30(d). Design output - Lack of or inadequate procedures Procedures for design output have not been established.
21 CFR 820.30(c). Design input - review and approval Design input requirements were not reviewed and approved by designated individual(s).
21 CFR 820.250(b). Sampling methods - Lack of or inadequate procedures Procedures to ensure sampling methods are adequate for their intended use have not been established.
21 CFR 820.22. Quality Audit/Reaudit - conducted Quality audits have not been performed.
21 CFR 820.72(a). Calibration, Inspection, etc. Procedures Lack of or Inadequ Procedures to ensure equipment is routinely calibrated have not been established.
21 CFR 820.140. Lack of or inadequate procedures for handling Procedures for product handling have not been established.
21 CFR 820.80(d). Lack of or inadequate final acceptance procedures Procedures for finished device acceptance have not been established.
21 CFR 820.198(a). Complaints Complaint files are not adequately maintained.
21 CFR 820.30(g). Design validation- Lack of or inadequate procedures Procedures for design validation have not been established.
Voluntary Action Indicated (VAI)
Warning letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports